Artículo

Development and characterisation of a confocal detection array for K-lines of heavy metals in big light matrix

Guarda, Jaime ; Valente, Mauro Andres Icon; Figueroa, Rodolfo

Otros responsables

Fecha

02/2021

Editorial

Pergamon-Elsevier Science Ltd

Revista

Radiation Physics and Chemistry (Oxford)

Título del libro

Nombre del evento

Fecha del evento

Institución organizadora

ISSN

0969-806X

eISSN

1879-0895

ISBN

Idioma

Inglés

Tipo del evento

Resumen

In this work we describe and simulate a setup which allows the X-ray Fluorescence (XRF) detection of gold nanoparticles (GNPs) excited by a convergent orthovoltage X-ray source, immersed in a big matrix of light elements (water). Under these conditions the problem is the poor Signal-to-Noise Ratio (SNR), which under normal irradiation conditions limit the detection of the GNPs. For the X-ray source (Teledyne CP120B) multiple calculations using SpekCalc software were run to determine an optimum external filtering thickness and collimation material, where was found that 1 mm of Cu external filtering improves the ratio and quantity of useful photons above the Au K-edge absorption energy. Monte Carlo (PENELOPE v2008) simulations were performed using the confocal geometry. The scenarios, including the composition (lead and bronze alloys), shape (cylindrical, conical) and location of the detectors were defined. Using a set of combinations for collimators in the X-ray source and detectors, where both are aligned to a unique focal point, it was possible to increase the SNR of the XRF signal with respect to Compton radiation in a specific region within the phantom, which makes possible to identify the location of the GNPs. The best SNR (185%) was achieved using a 4 cm length conical collimator made of lead. This would allow the development of a new technique for functional medical imaging, and according with the outcome of these simulations, in subsequent stages a prototype will be developed to validate the simulation outcomes, and confirm that it is possible to detect the location of GNPs from a human scale phantom based on their fluorescent emission.

Colecciones

Aprobación

Revisión

Complementado por

Referenciado por

Licencia Creative Commons

Excepto donde se indique lo contrario, la licencia de este ítem se describe como by-nc-sa