Mostrar el registro sencillo del ítem

dc.contributor.author
Kociubczyk, Alex Iván  
dc.contributor.author
Mendez, Claudia  
dc.contributor.author
Gregorutti, Ricardo  
dc.contributor.author
Ares, Alicia Esther  
dc.date.available
2016-12-19T14:51:21Z  
dc.date.issued
2015-04  
dc.identifier.citation
Kociubczyk, Alex Iván; Mendez, Claudia; Gregorutti, Ricardo; Ares, Alicia Esther; Electrochemical tests in stainless steel surgical implants; Elsevier; Procedia Materials Science; 9; 4-2015; 335-340  
dc.identifier.issn
2211-8128  
dc.identifier.uri
http://hdl.handle.net/11336/9700  
dc.description.abstract
In the present work the corrosion resistance of stainless steels was evaluated following the specifications of ASTM F-745 standards. The specimens were obtained by two methods, the conventional process of investment casting and countergravity low pressure casting. Cyclic polarization tests were performed according to ASTM F 2129, using a conventional three-electrode cell, with Pt as a counter electrode and saturated calomel as a reference electrode. The experiments were performed at (37 ± 1) °C and, before starting the test, the electrolyte was purged with N2 gas for 15 minutes. As electrolyte, a solution of 0.9% NaCl was used (pH≈7), the open circuit potential (OCP) was monitored for 1 hour and then the potentiodynamic scanning started in the anodic direction, at a speed of 0.167 mV/s, from a potential of 100 mV below the OCP. The sweep was reversed after passing the threshold of current density at a value hundred times greater than the current density of pitting. Measurements of electrochemical impedance spectroscopy were performed after one hour of exposure of the specimens to OCPs performed in a frequency range from 100 kHz to 1 mHz and potential amplitude of ± 10 mV. The microstructure of the samples was analyzed before and after testing using optical microscopy. The results indicated that the samples in general exhibit crack corrosion, and in some cases the occurrence of pitting corrosion was observed. Also, the capacity values were within the range corresponding to the electric double layer and the charge transfer resistance increased on the specimens obtained by conventional casting in ceramic mold.  
dc.format
application/pdf  
dc.language.iso
eng  
dc.publisher
Elsevier  
dc.rights
info:eu-repo/semantics/openAccess  
dc.rights.uri
https://creativecommons.org/licenses/by-nc-sa/2.5/ar/  
dc.source
International Congress of Science and Technology of Metallurgy and Materials, SAM – CONAMET 2014  
dc.subject
Biomaterials  
dc.subject
Corrosion  
dc.subject
Stainless Steel  
dc.subject.classification
Otras Ingeniería de los Materiales  
dc.subject.classification
Ingeniería de los Materiales  
dc.subject.classification
INGENIERÍAS Y TECNOLOGÍAS  
dc.title
Electrochemical tests in stainless steel surgical implants  
dc.type
info:eu-repo/semantics/article  
dc.type
info:ar-repo/semantics/artículo  
dc.type
info:eu-repo/semantics/publishedVersion  
dc.date.updated
2016-12-16T17:55:25Z  
dc.journal.volume
9  
dc.journal.pagination
335-340  
dc.journal.pais
Países Bajos  
dc.journal.ciudad
Amsterdam  
dc.description.fil
Fil: Kociubczyk, Alex Iván. Consejo Nacional de Investigaciones Cientificas y Tecnicas. Centro Cientifico Tecnológico Nordeste. Instituto de Materiales de Misiones; Argentina  
dc.description.fil
Fil: Mendez, Claudia. Consejo Nacional de Investigaciones Cientificas y Tecnicas. Centro Cientifico Tecnológico Nordeste. Instituto de Materiales de Misiones; Argentina  
dc.description.fil
Fil: Gregorutti, Ricardo. Provincia de Buenos Aires. Gobernacion. Comision de Investigaciones Cientificas. Laboratorio de Entrenamiento Multidisciplinario para la Investigación Tecnológica; Argentina  
dc.description.fil
Fil: Ares, Alicia Esther. Consejo Nacional de Investigaciones Cientificas y Tecnicas. Centro Cientifico Tecnológico Nordeste. Instituto de Materiales de Misiones; Argentina  
dc.journal.title
Procedia Materials Science  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/url/http://www.sciencedirect.com/science/article/pii/S2211812815001868  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/doi/http://dx.doi.org/10.1016/j.mspro.2015.05.002