dc.identifier.citation
Santibáñez, M.; Guillen, Y.; Chacón Obando, David; Figueroa, R.G.; Valente, Mauro Andres; Feasibility of dose enhancement assessment: Preliminary results by means of Gd-infused polymer gel dosimeter and Monte Carlo study; Pergamon-Elsevier Science Ltd; Applied Radiation and Isotopes; 141; 11-11-2018; 210-218
dc.identifier.issn
0969-8043
dc.identifier.uri
http://hdl.handle.net/11336/91583
dc.description.abstract
This work reports the experimental development of an integral Gd-infused dosimeter suitable for Gd dose enhancement assessment along with Monte Carlo simulations applied to determine the dose enhancement by radioactive and X-ray sources of interest in conventional and electronic brachytherapy. In this context, capability to elaborate a stable and reliable Gd-infused dosimeter was the first goal aimed at direct and accurate measurements of dose enhancement due to Gd presence. Dose-response was characterized for standard and Gd-infused PAGAT polymer gel dosimeters by means of optical transmission/absorbance. The developed Gd-infused PAGAT dosimeters demonstrated to be stable presenting similar dose-response as standard PAGAT within a linear trend up to 13 Gy along with good post-irradiation readout stability verified at 24 and 48 h. Additionally, dose enhancement was evaluated for Gd-infused PAGAT dosimeters by means of Monte Carlo (PENELOPE) simulations considering scenarios for isotopic and X-ray generator sources. The obtained results demonstrated the feasibility of obtaining a maximum enhancement around of (14 ± 1)% for 192 Ir source and an average enhancement of (70 ± 13)% for 241 Am. However, dose enhancement up to (267 ± 18)% may be achieved if suitable filtering is added to the 241 Am source. On the other hand, optimized X-ray spectra may attain dose enhancements up to (253 ± 22) %, which constitutes a promising future alternative for replacing radioactive sources by implementing electronic brachytherapy achieving high dose levels.
dc.format
application/pdf
dc.language.iso
eng
dc.publisher
Pergamon-Elsevier Science Ltd

dc.rights
info:eu-repo/semantics/openAccess
dc.rights.uri
https://creativecommons.org/licenses/by-nc-sa/2.5/ar/
dc.subject
ELECTRONIC BRACHYTHERAPY
dc.subject
GD DOSE ENHANCEMENT
dc.subject
PAGAT GEL DOSIMETRY
dc.subject.classification
Física Atómica, Molecular y Química

dc.subject.classification
Ciencias Físicas

dc.subject.classification
CIENCIAS NATURALES Y EXACTAS

dc.title
Feasibility of dose enhancement assessment: Preliminary results by means of Gd-infused polymer gel dosimeter and Monte Carlo study
dc.type
info:eu-repo/semantics/article
dc.type
info:ar-repo/semantics/artículo
dc.type
info:eu-repo/semantics/publishedVersion
dc.date.updated
2019-10-22T16:37:30Z
dc.journal.volume
141
dc.journal.pagination
210-218
dc.journal.pais
Estados Unidos

dc.description.fil
Fil: Santibáñez, M.. Universidad de La Frontera; Chile
dc.description.fil
Fil: Guillen, Y.. Universidad de La Frontera; Chile
dc.description.fil
Fil: Chacón Obando, David. Universidad Nacional. Departamento de Física; Costa Rica. Universidad Nacional de Córdoba. Facultad de Matemática, Astronomía y Física; Argentina
dc.description.fil
Fil: Figueroa, R.G.. Universidad de La Frontera; Chile
dc.description.fil
Fil: Valente, Mauro Andres. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto de Física Enrique Gaviola. Universidad Nacional de Córdoba. Instituto de Física Enrique Gaviola; Argentina. Universidad Nacional de Córdoba. Facultad de Matemática, Astronomía y Física; Argentina. Universidad de La Frontera; Chile
dc.journal.title
Applied Radiation and Isotopes

dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/url/https://www.sciencedirect.com/science/article/pii/S0969804317314197
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/doi/https://doi.org/10.1016/j.apradiso.2018.04.016