Evaluation of performance of an accelerator-based BNCT facility for the treatment of different tumor targets

Herrera, Maria SilviaIcon ; González, Sara JosefinaIcon ; Minsky, Daniel MauricioIcon ; Kreiner, Andres JuanIcon
Fecha de publicación: 09/2013
Editorial: Elsevier
Revista: Physica Medica
ISSN: 1120-1797
Idioma: Inglés
Tipo de recurso: Artículo publicado
Clasificación temática:
Otras Ciencias Físicas

Resumen

Purpose: Encouraging Boron Neutron Capture Therapy (BNCT) clinical results obtained in recent years have stimulated intense research to develop accelerator-based neutron sources to be installed in clinical facilities. In this work an assessment of an accelerator-based BNCT facility for the treatment of different tumor targets was performed, comparing the accelerator-derived results with reported reactor-based trials under similar conditions and subjected to the same clinical protocols. Materials and methods: A set of real image studies was used to cover clinical-like cases of brain and head-and-neck tumors. In addition, two clinical cases of malignant nodular melanoma treated at the RA-6 BNCT facility in Argentina were used to thoroughly compare the clinical dosimetry with the accelerator-derived results. Results: The minimum weighted dose delivered to the clinical target volume was higher than 30 Gy and 14 Gy for the brain tumor and head-and-neck cases, respectively, in agreement with those achieved in clinical applications. For the melanoma cases, the minimum tumor doses were equal or higher than those achieved with the RA-6 reactor for identical field orientation and protocol. The whole-body dose assessment showed that the maximum photon-equivalent doses for those normal organs close to the beam direction were below the upper limits considered in the protocols used in the present work. Conclusions: The obtained results indicate not only the good performance of the proposedbeam shaping assembly design associated to the facility but also the potential applicability of accelerator-based BNCT in the treatment of both superficial and deep-seated tumors.
Palabras clave: ACCELERATOR-BASED NEUTRON SOURCE , TREATMENT PLANNING ASSESSMENT , WHOLE-BODY DOSE ASSESSMENT
Ver el registro completo
 
Archivos asociados
Thumbnail
 
Tamaño: 1.197Mb
Formato: PDF
.
Descargar
Licencia
info:eu-repo/semantics/openAccess Excepto donde se diga explícitamente, este item se publica bajo la siguiente descripción: Creative Commons Attribution-NonCommercial-ShareAlike 2.5 Unported (CC BY-NC-SA 2.5)
Identificadores
URI: http://hdl.handle.net/11336/88235
DOI: http://dx.doi.org/10.1016/j.ejmp.2013.01.006
URL: https://www.sciencedirect.com/science/article/pii/S1120179713000070
Colecciones
Articulos(SEDE CENTRAL)
Articulos de SEDE CENTRAL
Citación
Herrera, Maria Silvia; González, Sara Josefina; Minsky, Daniel Mauricio; Kreiner, Andres Juan; Evaluation of performance of an accelerator-based BNCT facility for the treatment of different tumor targets; Elsevier; Physica Medica; 29; 5; 9-2013; 436-446
Compartir
Altmétricas
 

Enviar por e-mail
Separar cada destinatario (hasta 5) con punto y coma.
  • Facebook
  • X Conicet Digital
  • Instagram
  • YouTube
  • Sound Cloud
  • LinkedIn

Los contenidos del CONICET están licenciados bajo Creative Commons Reconocimiento 2.5 Argentina License

https://www.conicet.gov.ar/ - CONICET

Inicio

Explorar

  • Autores
  • Disciplinas
  • Comunidades

Estadísticas

Novedades

  • Noticias
  • Boletines

Ayuda

Acerca de

  • CONICET Digital
  • Equipo
  • Red Federal

Contacto

Godoy Cruz 2290 (C1425FQB) CABA – República Argentina – Tel: +5411 4899-5400 repositorio@conicet.gov.ar
TÉRMINOS Y CONDICIONES