Artículo
Estimation of viscoelastic arterial wall transfer function based on volume compensation method
Fecha de publicación:
12/2015
Editorial:
Springer
Revista:
Ifmbe Proceedings
ISSN:
1680-0737
Idioma:
Inglés
Tipo de recurso:
Artículo publicado
Clasificación temática:
Resumen
The viscoelastic arterial wall transfer function was determined noninvasively based on pressure and diameter signals of the finger artery measured sequentially using the volume compensation method. The uncompensated arterial diameter variations were converted to pressure using the estimated transfer function parameters. This pressure signal was negligible compared to measured pressure except at systole when reduction of 9 mmHg was achieved, thus correcting the volume compensation method’s error. To validate the method, central aortic pressure should be reconstructed from finger artery measurements and compared to invasive methods.
Archivos asociados
Licencia
Identificadores
Colecciones
Articulos(SEDE CENTRAL)
Articulos de SEDE CENTRAL
Articulos de SEDE CENTRAL
Citación
Casal, Leonardo; la Mura, Guillermo Mario; Estimation of viscoelastic arterial wall transfer function based on volume compensation method; Springer; Ifmbe Proceedings; 49; 12-2015; 23-26