Artículo
ICP-OES Determination of cobalt in natural water using a flow injection system after preconcentration on activated carbon
Farias, Gustavo Marcelo; Cerutti, Estela Soledad
; Gasquez, José Antonio; Olsina, Roberto Antonio
; Martinez, Luis Dante
Fecha de publicación:
11/2003
Editorial:
Perkin-elmer Corp
Revista:
Atomic Spectroscopy
ISSN:
0195-5373
e-ISSN:
2708-521X
Idioma:
Inglés
Tipo de recurso:
Artículo publicado
Clasificación temática:
Resumen
An on-line cobalt preconcentration system is described for the ICP-OES determination of cobalt in drinking water in a flow injection system. The methodology developed is simple and rapid. Trace amounts of cobalt were pre-concentrated by sorption on a conical mini column packed with activated carbon (AC) at pH 9.5, with retention of 95%. The cobalt was removed from the conical mini column with 20% (v/v) nitric acid. An enrichment factor of 95-fold for a sample volume of 50 mL was obtained. The detection limit for the preconcentration method proposed was found to be 20 ng L–1. The precision for 10 replicate determinations at the 5-µg L–1 Co level was 2.7% relative standard deviation, calculated from the peak heights obtained. The calibration graph preconcentration method for cobalt species was linear with a correlation coefficient of 0.9994 at levels near the detection limits up to at least 100 µg L–1. The method was successfully applied to the determination of cobalt in natural water samples.
Archivos asociados
Licencia
Identificadores
Colecciones
Articulos(CCT - SAN LUIS)
Articulos de CTRO.CIENTIFICO TECNOL.CONICET - SAN LUIS
Articulos de CTRO.CIENTIFICO TECNOL.CONICET - SAN LUIS
Articulos(INQUISAL)
Articulos de INST. DE QUIMICA DE SAN LUIS
Articulos de INST. DE QUIMICA DE SAN LUIS
Citación
Farias, Gustavo Marcelo; Cerutti, Estela Soledad; Gasquez, José Antonio; Olsina, Roberto Antonio; Martinez, Luis Dante; ICP-OES Determination of cobalt in natural water using a flow injection system after preconcentration on activated carbon; Perkin-elmer Corp; Atomic Spectroscopy; 24; 6; 11-2003; 213-217
Compartir
Altmétricas