An on-line mercury pre-concentration and analysis system using atomic absorption spectrometry (AAS) and flow injection (FI) with cold vapor (CV) generation was studied. The mercury was retained as a mercury-2-(5-bromo-2-pyridylazo)-5-diethplaminophenol [Hg(II)-(5-Br-PADAP)] complex, at pH 9.2. The mercury complex was removed from the knotted reactor (KR) with 3 mol/L hydrochloric acid. An enrichment factor of 25 was obtained with respect to CV AAS. The detection limit for the pre concentration of 25 ml, aqueous solution was 5 ng/L. The precision, calculated at peak height for 10 replicate determinations at the 1 µg/L-Hg level, was 2.8% relative standard deviation (RSD): The calibration curve using;. the pre-concentration system for mercury was linear with a correlation coefficient of 0.9995 at levels near the detection limits up to at least 100 µg/L. The method was successfully applied to the determination of mercury in drinking water samples.
Archivos asociados
Licencia
Identificadores
Colecciones
Articulos(INQUISAL)
Articulos de INST. DE QUIMICA DE SAN LUIS
Articulos de INST. DE QUIMICA DE SAN LUIS
Citación
Altamirano, Jorgelina Cecilia; Wuilloud, Rodolfo German; Gasquez, José Antonio; Olsina, Roberto Antonio; Martinez, Luis Dante; On-line adsorption/preconcentration using a knotted reactor for total Hg determination in drinking water by flow injection CV-AAS; Perkin-Elmer Corporation; Atomic Spectroscopy; 22; 5; 5-2001; 398-404
Compartir