Mostrar el registro sencillo del ítem

dc.contributor.author
Wuilloud, Rodolfo German  
dc.contributor.author
Acevedo, H.A.  
dc.contributor.author
Vazquez, F.A.  
dc.contributor.author
Martinez, Luis Dante  
dc.date.available
2020-03-17T21:12:23Z  
dc.date.issued
2002-10  
dc.identifier.citation
Wuilloud, Rodolfo German; Acevedo, H.A.; Vazquez, F.A.; Martinez, Luis Dante; Determination of lead in drinking water by ICP-AES with ultrasonic nebulization and flow-injection on-line preconcentration using an amberlite XAD-16 resin; Taylor & Francis; Analytical Letters; 35; 10; 10-2002; 1649-1665  
dc.identifier.issn
0003-2719  
dc.identifier.uri
http://hdl.handle.net/11336/99943  
dc.description.abstract
An on-line lead preconcentration and determination system implemented with inductively coupled plasma atomic emission spectrometry (ICP-AES) associated to flow injection (FI) with ultrasonic nebulization (USN) was studied. For the retention of lead, 2-(5-bromo-2-pyridylazo)-5-diethyl-aminophenol (5-Br-PADAP) and Amberlite XAD-16 were used, at pH 9.2. The lead was removed from the microcolumn in counter flow with nitric acid. A total enhancement factor of 150 was obtained with respect to ICP-AES using pneumatic nebulization (15.0 for USN and 10.0 for column). The detection limit value for the preconcentration of 10 mL of aqueous solution of Pb was 0.2 ng mL-1. The precision for 10 replicate determinations at the 5 ng mL-1 Pb levels was 2.8% relative standard deviation (RSD), calculated with the peak heights obtained. The calibration graph using the preconcentration system for lead was linear with a correlation coefficient of 0.9997 at levels near the detection limits up to at least 100 ng mL-1. The method was successfully applied to the determination of lead in drinking water samples.  
dc.format
application/pdf  
dc.language.iso
eng  
dc.publisher
Taylor & Francis  
dc.rights
info:eu-repo/semantics/openAccess  
dc.rights.uri
https://creativecommons.org/licenses/by-nc-sa/2.5/ar/  
dc.subject
5-BR-PADAP  
dc.subject
FLOW INJECTION  
dc.subject
ICP-AES  
dc.subject
LEAD IN DRINKING WATER SAMPLES  
dc.subject
PRECONCENTRATION AND DETERMINATION  
dc.subject.classification
Química Analítica  
dc.subject.classification
Ciencias Químicas  
dc.subject.classification
CIENCIAS NATURALES Y EXACTAS  
dc.title
Determination of lead in drinking water by ICP-AES with ultrasonic nebulization and flow-injection on-line preconcentration using an amberlite XAD-16 resin  
dc.type
info:eu-repo/semantics/article  
dc.type
info:ar-repo/semantics/artículo  
dc.type
info:eu-repo/semantics/publishedVersion  
dc.date.updated
2020-02-13T20:05:24Z  
dc.journal.volume
35  
dc.journal.number
10  
dc.journal.pagination
1649-1665  
dc.journal.pais
Estados Unidos  
dc.description.fil
Fil: Wuilloud, Rodolfo German. Universidad Nacional de San Luis. Facultad de Química, Bioquímica y Farmacia. Departamento de Química. Área de Química Analítica; Argentina. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - San Luis. Instituto de Química de San Luis. Universidad Nacional de San Luis. Facultad de Química, Bioquímica y Farmacia. Instituto de Química de San Luis; Argentina  
dc.description.fil
Fil: Acevedo, H.A.. Universidad Nacional de San Luis. Facultad de Química, Bioquímica y Farmacia. Departamento de Química. Área de Química Analítica; Argentina  
dc.description.fil
Fil: Vazquez, F.A.. Universidad Nacional de San Luis. Facultad de Química, Bioquímica y Farmacia. Departamento de Química. Área de Química Analítica; Argentina  
dc.description.fil
Fil: Martinez, Luis Dante. Universidad Nacional de San Luis. Facultad de Química, Bioquímica y Farmacia. Departamento de Química. Área de Química Analítica; Argentina. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - San Luis. Instituto de Química de San Luis. Universidad Nacional de San Luis. Facultad de Química, Bioquímica y Farmacia. Instituto de Química de San Luis; Argentina  
dc.journal.title
Analytical Letters  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/doi/https://doi.org/10.1081/AL-120013045  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/url/https://www.tandfonline.com/doi/abs/10.1081/AL-120013045