Artículo
A sensitive spectrophotometric method for lead determination by flow injection analysis with on-line preconcentration
Fecha de publicación:
12/2004
Editorial:
Elsevier Science
Revista:
Talanta
ISSN:
0039-9140
Idioma:
Inglés
Tipo de recurso:
Artículo publicado
Clasificación temática:
Resumen
A new flow injection (FI) system for the determination of Pb(II) at trace level with a preconcentration step and spectrophotometric detection is proposed. It is based on preconcentration of lead ions on chitosan and dithizone-lead complex formation in aqueous medium (pH 9). The chemicals and FIA variables influencing the performance of the system were optimized and applied to the determination of lead in natural, well, and drinking water samples. It is a simple, highly sensitive, and low cost alternative methodology. The method provided a linear rage between 25 and 250 μg l-1, a detection limit of 5.0 ng ml-1 and a sample throughput of 15 h-1. The obtained results of spiked samples are in good agreement between the proposed method and ICP-AES.
Palabras clave:
FIA
,
LEAD
,
PRECONCENTRATION
,
SPECTROPHOTOMETRIC
Archivos asociados
Licencia
Identificadores
Colecciones
Articulos(INQUISUR)
Articulos de INST.DE QUIMICA DEL SUR
Articulos de INST.DE QUIMICA DEL SUR
Citación
Di Nezio, Maria Susana; Palomeque, Miriam Edid; Fernández Band, Beatriz Susana; A sensitive spectrophotometric method for lead determination by flow injection analysis with on-line preconcentration; Elsevier Science; Talanta; 63; 2; 12-2004; 405-409
Compartir
Altmétricas