Narata, Ana Paula; Blasco, Jordi; Roman, Luis San; Macho, Juan Miguel; Fernandez, Hector; et al.; Early Results in Flow Diverter Sizing by Computational Simulation: Quantification of Size Change and Simulation Error Assessment; Oxford University Press; Operative Neurosurgery; 15; 5; 11-2018; 557-566  
dc.identifier.issn
2332-4252  
dc.identifier.uri
http://hdl.handle.net/11336/95908  
dc.description.abstract
BACKGROUND: Sizing of flow diverters (FDs) stent in the treatment of intracranial aneurysms is a challenging task due to the change of stent length after implantation. OBJECTIVE: To quantify the size change and assess the error in length prediction in 82 simulated FD deployments. METHODS: Eighty-two consecutive patients treated with FDs were retrospectively analyzed. Implanted FD length was measured from angiographic images and compared to the nominal sizes of the implanted device. Length change was obtained by subtracting the nominal length from the real length and dividing by the nominal length. Implanted devices were simulated on 3-dimensional models of each patient. Simulation error was obtained by subtracting real length from simulated length and dividing by the real length of the FD. Subanalysis was done using ANOVA. Statistical significance was set to P < .05, and bootstrap resampling was used. RESULTS: When assessing the length change of the FD after implantation, changes of 30% in average and up to 80% with reference to the nominal length of the device were observed. The simulation results showed a lower error of 3.52% in averagewith a maximum of 30%. Paired t-test showed nonsignificant differences between measured and real length (P=.07, with the mean of differences at 0.45 mm, 95% confidence interval [-0.950 0.038]). CONCLUSION: Nominal length is not an accurate sizing metric when choosing the size of an FD irrespective of the brand and manufacturer. Good estimation of the final length of the stent after deployment as expressed by an error of 3.5% in average.  
dc.format
application/pdf  
dc.language.iso
eng  
dc.publisher
Oxford University Press  
dc.rights
info:eu-repo/semantics/openAccess  
dc.rights.uri
https://creativecommons.org/licenses/by-nc-sa/2.5/ar/  
dc.subject
angiogram  
dc.subject
stents  
dc.subject
intracranial aneurysm  
dc.subject
devices  
dc.subject
medical  
dc.subject
military deployment  
dc.subject
fluid flow  
dc.subject.classification
Ciencias de la Información y Bioinformática  
dc.subject.classification
Ciencias de la Computación e Información  
dc.subject.classification
CIENCIAS NATURALES Y EXACTAS  
dc.title
Early Results in Flow Diverter Sizing by Computational Simulation: Quantification of Size Change and Simulation Error Assessment  
dc.type
info:eu-repo/semantics/article  
dc.type
info:ar-repo/semantics/artículo  
dc.type
info:eu-repo/semantics/publishedVersion  
dc.date.updated
2019-10-22T18:01:13Z  
dc.journal.volume
15  
dc.journal.number
5  
dc.journal.pagination
557-566  
dc.journal.pais
Reino Unido  
dc.description.fil
Fil: Narata, Ana Paula. Inserm; Francia. Universite de Tours; Francia  
dc.description.fil
Fil: Blasco, Jordi. Hospital Clinic Barcelona; España  
dc.description.fil
Fil: Roman, Luis San. Hospital Clinic Barcelona; España  
dc.description.fil
Fil: Macho, Juan Miguel. Hospital Clinic Barcelona; España  
dc.description.fil
Fil: Fernandez, Hector. Galgo Medical Sl; España  
dc.description.fil
Fil: Moyano, Raquel Kale. Galgo Medical Sl; España  
dc.description.fil
Fil: Winzenrieth, Renaud. Galgo Medical Sl; España  
dc.description.fil
Fil: Larrabide, Ignacio. Universidad Nacional del Centro de la Provincia de Buenos Aires; Argentina. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina  
dc.journal.title
Operative Neurosurgery  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/url/https://academic.oup.com/ons/article-abstract/15/5/557/4812488  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/doi/http://dx.doi.org/10.1093/ons/opx288