Mostrar el registro sencillo del ítem
dc.contributor.author
Llamas, Natalia Elizabeth
dc.contributor.author
Di Nezio, Maria Susana
dc.contributor.author
Palomeque, Miriam Edid
dc.contributor.author
Fernández Band, Beatriz Susana
dc.date.available
2020-01-10T21:16:08Z
dc.date.issued
2005-05-10
dc.identifier.citation
Llamas, Natalia Elizabeth; Di Nezio, Maria Susana; Palomeque, Miriam Edid; Fernández Band, Beatriz Susana; Automated turbidimetric determination of cyclamate in low calorie soft drinks and sweeteners without pre-treatment; Elsevier Science; Analytica Chimica Acta; 539; 1-2; 10-5-2005; 301-304
dc.identifier.issn
0003-2670
dc.identifier.uri
http://hdl.handle.net/11336/94422
dc.description.abstract
A flow injection turbidimetric method for determination of cyclamate in low calorie soft drinks and sweeteners without pre-treatment is proposed. It was based on the oxidation of the sulphamic group of cyclamate, to sulphate by addition of nitrite. Then, a precipitate of sulphate was obtained by reaction with barium chloride, in presence of poly(vinyl alcohol) (PVA) in perchloric acid solution, at 30°C. The analytical signal was measured at 420 nm. The method presented a linear range between 0.015 and 0.120% (w/v) and the calibration graph was S = (7.792 ± 0.179) X - (0.114 ± 0.012), R2 = 0.999. A reproducibility of 5.9% was obtained from nine calibration graphs that were carried out on different days and with different conditions (standard solution, reagents solution, etc.). The detection and quantification limits were 0.006 and 0.02% (w/v), respectively, the sample throughput 45 h-1 (by considering the wash cycle and the time consuming to eliminate the clean solution from the system). The obtained results were agreement with that obtained with the standard method.
dc.format
application/pdf
dc.language.iso
eng
dc.publisher
Elsevier Science
dc.rights
info:eu-repo/semantics/openAccess
dc.rights.uri
https://creativecommons.org/licenses/by-nc-nd/2.5/ar/
dc.subject
CYCLAMATE
dc.subject
FIA
dc.subject
SWEETENERS
dc.subject
TURBIDIMETRIC
dc.subject.classification
Química Analítica
dc.subject.classification
Ciencias Químicas
dc.subject.classification
CIENCIAS NATURALES Y EXACTAS
dc.title
Automated turbidimetric determination of cyclamate in low calorie soft drinks and sweeteners without pre-treatment
dc.type
info:eu-repo/semantics/article
dc.type
info:ar-repo/semantics/artículo
dc.type
info:eu-repo/semantics/publishedVersion
dc.date.updated
2019-11-27T14:52:36Z
dc.journal.volume
539
dc.journal.number
1-2
dc.journal.pagination
301-304
dc.journal.pais
Países Bajos
dc.journal.ciudad
Amsterdam
dc.description.fil
Fil: Llamas, Natalia Elizabeth. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Bahía Blanca. Instituto de Química del Sur. Universidad Nacional del Sur. Departamento de Química. Instituto de Química del Sur; Argentina
dc.description.fil
Fil: Di Nezio, Maria Susana. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Bahía Blanca. Instituto de Química del Sur. Universidad Nacional del Sur. Departamento de Química. Instituto de Química del Sur; Argentina
dc.description.fil
Fil: Fernández Band, Beatriz Susana. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Bahía Blanca. Instituto de Química del Sur. Universidad Nacional del Sur. Departamento de Química. Instituto de Química del Sur; Argentina
dc.journal.title
Analytica Chimica Acta
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/url/https://www.sciencedirect.com/science/article/pii/S0003267005004964
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/doi/http://dx.doi.org/10.1016/j.aca.2005.03.044