Mostrar el registro sencillo del ítem
dc.contributor.author
Perez, Roberto Daniel

dc.contributor.author
Falchini, Gisele Evangelina

dc.contributor.author
Cometto Vincente, Franco

dc.contributor.author
Soares, L.
dc.contributor.author
Poletti, M.E.
dc.contributor.author
Sanchez, Hector Jorge

dc.date.available
2019-12-05T21:52:37Z
dc.date.issued
2019-02-01
dc.identifier.citation
Perez, Roberto Daniel; Falchini, Gisele Evangelina; Cometto Vincente, Franco; Soares, L.; Poletti, M.E.; et al.; A new XRF spectrometer using a crystal monochromator and parallel plates beam guides; Elsevier Science; Nuclear Instruments and Methods in Physics Research B: Beam Interactions with Materials and Atoms; 440; 1-2-2019; 48-53
dc.identifier.issn
0168-583X
dc.identifier.uri
http://hdl.handle.net/11336/91563
dc.description.abstract
Conventional x-ray fluorescence analysis (XRF) is a powerful instrumental technique for elemental quantification of samples with a wide range of matrices coming from industry, medicine, biology, environmental and archaeometry. The scattering of the x-ray tube excitation beam from the matrix is mainly responsible of the strong background in XRF spectra which restricts the detection limit (DL) to a few ppm (w/w). One possible strategy to reduce the DL is the monochromatization of the excitation beam with a crystal monochromator. In this case, the optimization of the system components is strongly required to obtain the highest excitation count rate over the sample, since it is considerable reduced by the short energy band pass of the monochromator. In the present work an optical device composed by a Si(100) crystal monochromator and two parallel plates beam guides (PPBG) was characterized and including in a conventional XRF spectrometer to improve the DLs. The PPBGs are composed by several thin glass plate reflectors placed together and separated by spacers. Inside the gap, x-rays are transmitted by total reflection contributing to the photon flux at the output of the optical device. The first PPBG reduces the divergence of the beam striking the crystal, and the second one collects the monochromatic beam that emerges from the crystal. In this way, the design of the optical device is quite compact keeping a high photon flux at the output thanks to the wide acceptance solid angle of the PPBGs. The new XRF spectrometer was applied to the analysis of biological tissues. The first results obtained in this field are shown in the present work where the performance of the spectrometer can be evaluated.
dc.format
application/pdf
dc.language.iso
eng
dc.publisher
Elsevier Science

dc.rights
info:eu-repo/semantics/restrictedAccess
dc.rights.uri
https://creativecommons.org/licenses/by-nc-sa/2.5/ar/
dc.subject
BEAM GUIDES
dc.subject
CRYSTAL MONOCHROMATOR
dc.subject
PARALLEL PLATES
dc.subject
TRACE ANALYSIS
dc.subject
XRF
dc.subject.classification
Física Atómica, Molecular y Química

dc.subject.classification
Ciencias Físicas

dc.subject.classification
CIENCIAS NATURALES Y EXACTAS

dc.title
A new XRF spectrometer using a crystal monochromator and parallel plates beam guides
dc.type
info:eu-repo/semantics/article
dc.type
info:ar-repo/semantics/artículo
dc.type
info:eu-repo/semantics/publishedVersion
dc.date.updated
2019-10-22T16:31:48Z
dc.journal.volume
440
dc.journal.pagination
48-53
dc.journal.pais
Países Bajos

dc.journal.ciudad
Amsterdam
dc.description.fil
Fil: Perez, Roberto Daniel. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto de Física Enrique Gaviola. Universidad Nacional de Córdoba. Instituto de Física Enrique Gaviola; Argentina. Universidad Nacional de Córdoba. Facultad de Matemática, Astronomía y Física; Argentina
dc.description.fil
Fil: Falchini, Gisele Evangelina. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto de Física Enrique Gaviola. Universidad Nacional de Córdoba. Instituto de Física Enrique Gaviola; Argentina. Universidad Nacional de Córdoba. Facultad de Matemática, Astronomía y Física; Argentina
dc.description.fil
Fil: Cometto Vincente, Franco. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto de Física Enrique Gaviola. Universidad Nacional de Córdoba. Instituto de Física Enrique Gaviola; Argentina. Universidad Nacional de Córdoba. Facultad de Matemática, Astronomía y Física; Argentina
dc.description.fil
Fil: Soares, L.. Universidade de Sao Paulo; Brasil
dc.description.fil
Fil: Poletti, M.E.. Universidade de Sao Paulo; Brasil
dc.description.fil
Fil: Sanchez, Hector Jorge. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto de Física Enrique Gaviola. Universidad Nacional de Córdoba. Instituto de Física Enrique Gaviola; Argentina. Universidad Nacional de Córdoba. Facultad de Matemática, Astronomía y Física; Argentina
dc.journal.title
Nuclear Instruments and Methods in Physics Research B: Beam Interactions with Materials and Atoms

dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/url/https://www.sciencedirect.com/science/article/pii/S0168583X18306918
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/doi/https://doi.org/10.1016/j.nimb.2018.11.030
Archivos asociados