Mostrar el registro sencillo del ítem

dc.contributor.author
Schuster, Jonathan Maximiliano  
dc.contributor.author
Schvezov, Carlos Enrique  
dc.contributor.author
Rosenberger, Mario Roberto  
dc.date.available
2019-10-23T18:17:20Z  
dc.date.issued
2018-12  
dc.identifier.citation
Schuster, Jonathan Maximiliano; Schvezov, Carlos Enrique; Rosenberger, Mario Roberto; Construction and calibration of a goniometer to measure contact angles and calculate the surface free energy in solids with uncertainty analysis; Elsevier; International Journal of Adhesion and Adhesives; 87; 12-2018; 205-215  
dc.identifier.issn
0143-7496  
dc.identifier.uri
http://hdl.handle.net/11336/87114  
dc.description.abstract
Here, we present the construction and calibration of a low-cost goniometer to measure contact angles by the sessile drop method. Besides, we propose a simple and fast method to calculate the uncertainty in the determination of the surface free energy (SFE) and its polar and dispersive components through the Owens-Wendt model and tested it by using two testing liquids. The goniometer performance and the SFE uncertainty were determined on two polymers: polytetrafluorethylene (PTFE) and polyoxymethylene (POM), by using water and methylene iodide. The values of contact angle measured were used to calculate the SFE and its components with their errors. The SFE values obtained for PTFE were 17.57–17.91 mJ/m2, with a relative error lower than 5.5%, whereas those for POM were 42.80–43.23 mJ/m2, with a relative error lower than 4.3%. Both the SFE values and the errors were in the range of those previously reported. Based on the mathematical analysis of the uncertainty propagation in the determination of SFE, we concluded that the uncertainty is minimized when the testing liquids are an apolar liquid and water.  
dc.format
application/pdf  
dc.language.iso
eng  
dc.publisher
Elsevier  
dc.rights
info:eu-repo/semantics/openAccess  
dc.rights.uri
https://creativecommons.org/licenses/by-nc-nd/2.5/ar/  
dc.subject
CALIBRATION  
dc.subject
EXPERIMENTAL ERRORS  
dc.subject
GONIOMETER  
dc.subject
SESSILE DROP  
dc.subject
SURFACE FREE ENERGY  
dc.subject.classification
Ingeniería de los Materiales  
dc.subject.classification
Ingeniería de los Materiales  
dc.subject.classification
INGENIERÍAS Y TECNOLOGÍAS  
dc.title
Construction and calibration of a goniometer to measure contact angles and calculate the surface free energy in solids with uncertainty analysis  
dc.type
info:eu-repo/semantics/article  
dc.type
info:ar-repo/semantics/artículo  
dc.type
info:eu-repo/semantics/publishedVersion  
dc.date.updated
2019-10-17T14:53:51Z  
dc.identifier.eissn
0143-7496  
dc.journal.volume
87  
dc.journal.pagination
205-215  
dc.journal.pais
Países Bajos  
dc.journal.ciudad
Amsterdam  
dc.description.fil
Fil: Schuster, Jonathan Maximiliano. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Nordeste. Instituto de Materiales de Misiones. Universidad Nacional de Misiones. Facultad de Ciencias Exactas Químicas y Naturales. Instituto de Materiales de Misiones; Argentina  
dc.description.fil
Fil: Schvezov, Carlos Enrique. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Nordeste. Instituto de Materiales de Misiones. Universidad Nacional de Misiones. Facultad de Ciencias Exactas Químicas y Naturales. Instituto de Materiales de Misiones; Argentina  
dc.description.fil
Fil: Rosenberger, Mario Roberto. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Nordeste. Instituto de Materiales de Misiones. Universidad Nacional de Misiones. Facultad de Ciencias Exactas Químicas y Naturales. Instituto de Materiales de Misiones; Argentina  
dc.journal.title
International Journal of Adhesion and Adhesives  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/doi/http://dx.doi.org/10.1016/j.ijadhadh.2018.10.012  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/url/https://www.sciencedirect.com/science/article/pii/S0143749618302422