Mostrar el registro sencillo del ítem
dc.contributor.author
Guzmán, Andrés Fernando
dc.contributor.author
Arciniegas Castilla, Alejandro
dc.contributor.author
Guarnieri, Fabio Ariel
dc.contributor.author
Ramírez Rodríguez, Fernando
dc.date.available
2016-12-01T17:10:49Z
dc.date.issued
2013-09
dc.identifier.citation
Guzmán, Andrés Fernando; Arciniegas Castilla, Alejandro ; Guarnieri, Fabio Ariel; Ramírez Rodríguez, Fernando; Intraocular Pressure: Goldmann Tonometry, Computational Model, and Calibration Equation; Lippincott Williams; Journal Of Glaucoma.; 22; 1; 9-2013; 10-14
dc.identifier.issn
1057-0829
dc.identifier.uri
http://hdl.handle.net/11336/8580
dc.description.abstract
It has been shown for some time that several visual pathologies are the result of special conditions of geometry (ECR: External Curvature Radius, CCT: Central Corneal Thickness), mechanical properties (E: Young's modulus), and intraocular pressure (IOP). There is one weakness on diagnosis of these pathologies related to the determining of true IOP. On this measure, it is questioning its validity with any type of tonometer and especially with Goldmann Applanation Tonometer (GAT), because it is calibrated for adult healthy eyes. In this work, the calibration of the measurements of the IOP through GAT is done using a simulation of mechanical corneal behaviour through the applanation process using the method of the finite element modeling. This methodology is follow for establishing the range of real IOP readings of patients to obtain a reliable measure of pressure. From statistical variable analysis point of view, supported on computational modelling, a calibration equation is established for the device that relates the true IOP with the ECR, ECC and GAT measurement, subject to a range of application.
dc.format
application/pdf
dc.language.iso
eng
dc.publisher
Lippincott Williams
dc.rights
info:eu-repo/semantics/openAccess
dc.rights.uri
https://creativecommons.org/licenses/by-nc-sa/2.5/ar/
dc.subject
Tonometry
dc.subject
Intraocular Pressure
dc.subject
Goldmanns Tonometer
dc.subject
Finite Element Method
dc.subject.classification
Biotecnología relacionada con la Salud
dc.subject.classification
Biotecnología de la Salud
dc.subject.classification
CIENCIAS MÉDICAS Y DE LA SALUD
dc.title
Intraocular Pressure: Goldmann Tonometry, Computational Model, and Calibration Equation
dc.type
info:eu-repo/semantics/article
dc.type
info:ar-repo/semantics/artículo
dc.type
info:eu-repo/semantics/publishedVersion
dc.date.updated
2016-11-24T14:13:26Z
dc.journal.volume
22
dc.journal.number
1
dc.journal.pagination
10-14
dc.journal.pais
Estados Unidos
dc.journal.ciudad
Philadelphia
dc.description.fil
Fil: Guzmán, Andrés Fernando. Universidad del Norte, Barranquilla; Colombia
dc.description.fil
Fil: Arciniegas Castilla, Alejandro . Instituto Barraquer de América, Bogotá; Colombia
dc.description.fil
Fil: Guarnieri, Fabio Ariel. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico. Centro de Investigación de Métodos Computacionales; Argentina
dc.description.fil
Fil: Ramírez Rodríguez, Fernando. Universidad de los Andes, Bogotá; Colombia