dc.contributor.author
Auzmendi, Jerónimo Andrés  
dc.contributor.author
Moffatt, Luciano  
dc.date.available
2019-03-11T20:03:46Z  
dc.date.issued
2010-01  
dc.identifier.citation
Auzmendi, Jerónimo Andrés; Moffatt, Luciano; Increasing the reliability of solution exchanges by monitoring solenoid valve actuation; Elsevier Science; Journal of Neuroscience Methods; 185; 2; 1-2010; 280-283  
dc.identifier.issn
0165-0270  
dc.identifier.uri
http://hdl.handle.net/11336/71390  
dc.description.abstract
Solenoid valves are a core component of most solution perfusion systems used in neuroscience research. As they open and close, they control the flow of solution through each perfusion line, thereby modulating the timing and sequence of chemical stimulation. The valves feature a ferromagnetic plunger that moves due to the magnetization of the solenoid and returns to its initial position with the aid of a spring. The delays between the time of voltage application or removal and the actual opening or closing of the valve are difficult to predict beforehand and have to be measured experimentally. Here we propose a simple method for monitoring whether and when the solenoid valve opens and closes. The proposed method detects the movement of the plunger as it generates a measurable signal on the solenoid that surrounds it. Using this plunger signal, we detected the opening and closing of diaphragm and pinch solenoid valves with a systematic error of less than 2 ms. After this systematic error is subtracted, the trial-to-trial error was below 0.2 ms. © 2009 Elsevier B.V. All rights reserved.  
dc.format
application/pdf  
dc.language.iso
eng  
dc.publisher
Elsevier Science  
dc.rights
info:eu-repo/semantics/openAccess  
dc.rights.uri
https://creativecommons.org/licenses/by-nc-sa/2.5/ar/  
dc.subject
Concentration Jump  
dc.subject
Patch Clamp  
dc.subject
Perfusion  
dc.subject
Receptor Kinetics  
dc.subject
Solenoid Valve  
dc.subject.classification
Otras Ciencias Biológicas  
dc.subject.classification
Ciencias Biológicas  
dc.subject.classification
CIENCIAS NATURALES Y EXACTAS  
dc.title
Increasing the reliability of solution exchanges by monitoring solenoid valve actuation  
dc.type
info:eu-repo/semantics/article  
dc.type
info:ar-repo/semantics/artículo  
dc.type
info:eu-repo/semantics/publishedVersion  
dc.date.updated
2019-03-08T20:29:49Z  
dc.journal.volume
185  
dc.journal.number
2  
dc.journal.pagination
280-283  
dc.journal.pais
Países Bajos  
dc.journal.ciudad
Amsterdam  
dc.description.fil
Fil: Auzmendi, Jerónimo Andrés. Consejo Nacional de Investigaciones Científicas y Técnicas. Oficina de Coordinación Administrativa Ciudad Universitaria. Instituto de Química, Física de los Materiales, Medioambiente y Energía. Universidad de Buenos Aires. Facultad de Ciencias Exactas y Naturales. Instituto de Química, Física de los Materiales, Medioambiente y Energía; Argentina  
dc.description.fil
Fil: Moffatt, Luciano. Consejo Nacional de Investigaciones Científicas y Técnicas. Oficina de Coordinación Administrativa Ciudad Universitaria. Instituto de Química, Física de los Materiales, Medioambiente y Energía. Universidad de Buenos Aires. Facultad de Ciencias Exactas y Naturales. Instituto de Química, Física de los Materiales, Medioambiente y Energía; Argentina  
dc.journal.title
Journal of Neuroscience Methods  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/url/https://www.sciencedirect.com/science/article/pii/S0165027009005378  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/doi/https://doi.org/10.1016/j.jneumeth.2009.10.002