Artículo
Assessment of thermal effects in a model of the human head implanted with a wireless active microvalve for the treatment of glaucoma creating a filtering bleb
Fecha de publicación:
04/2017
Editorial:
IOP Publishing
Revista:
Physics In Medicine And Biology
ISSN:
0031-9155
Idioma:
Inglés
Tipo de recurso:
Artículo publicado
Clasificación temática:
Resumen
A 3D anatomical computational model is developed to assess thermal effects due to exposure to the electromagnetic field required to power a new investigational active implantable microvalve for the treatment of glaucoma. Such a device, located in the temporal superior eye quadrant, produces a filtering bleb, which is included in the geometry of the model, together with the relevant ocular structures. The electromagnetic field source?a planar coil?as well as the microvalve antenna and casing are also included. Exposure to the electromagnetic field source of an implanted and a non-implanted subject are simulated by solving a magnetic potential formulation, using the finite element method. The maximum SAR10 is reached in the eyebrow and remains within the limits suggested by the IEEE and ICNIRP standards. The anterior chamber, filtering bleb, iris and ciliary body are the ocular structures where more absorption occurs. The temperature rise distribution is also obtained by solving the bioheat equation with the finite element method. The numerical results are compared with the in vivo measurements obtained from four rabbits implanted with the microvalve and exposed to the electromagnetic field source.
Palabras clave:
Microvalve
,
Thermal Model
,
Electromagnetic Exposition
,
Glaucoma
Archivos asociados
Licencia
Identificadores
Colecciones
Articulos(CIMEC)
Articulos de CENTRO DE INVESTIGACION DE METODOS COMPUTACIONALES
Articulos de CENTRO DE INVESTIGACION DE METODOS COMPUTACIONALES
Articulos(INTEC)
Articulos de INST.DE DES.TECNOL.PARA LA IND.QUIMICA (I)
Articulos de INST.DE DES.TECNOL.PARA LA IND.QUIMICA (I)
Citación
Schaumburg, Federico; Guarnieri, Fabio Ariel; Assessment of thermal effects in a model of the human head implanted with a wireless active microvalve for the treatment of glaucoma creating a filtering bleb; IOP Publishing; Physics In Medicine And Biology; 62; 9; 4-2017; 191-203
Compartir
Altmétricas