Repositorio Institucional
Repositorio Institucional
CONICET Digital
  • Inicio
  • EXPLORAR
    • AUTORES
    • DISCIPLINAS
    • COMUNIDADES
  • Estadísticas
  • Novedades
    • Noticias
    • Boletines
  • Ayuda
    • General
    • Datos de investigación
  • Acerca de
    • CONICET Digital
    • Equipo
    • Red Federal
  • Contacto
JavaScript is disabled for your browser. Some features of this site may not work without it.
  • INFORMACIÓN GENERAL
  • RESUMEN
  • ESTADISTICAS
 
Artículo

Asymmetry of characteristic X-ray peaks obtained by a Si(Li) detector

Visñovezky, Claudia Analía; Limandri, Silvina PaolaIcon ; Canafoglia, Maria Elena; Bonetto, Rita DomingaIcon ; Trincavelli, Jorge CarlosIcon
Fecha de publicación: 05/2007
Editorial: Pergamon-Elsevier Science Ltd
Revista: Spectrochimica Acta Part B: Atomic Spectroscopy
ISSN: 0584-8547
Idioma: Inglés
Tipo de recurso: Artículo publicado
Clasificación temática:
Astronomía

Resumen

The asymmetry of the characteristic X-ray peaks obtained using a Si(Li) detector is mainly due to incomplete charge collection. Impurities and defects in the crystalline structure of Si can act as "traps" for holes and electrons in their trip toward the detector electrodes. Therefore, the collected charge, and consequently the detected energy, is smaller than the expected one. The global effect is that peaks may present a "tail" toward the low energy side. This phenomenon is more important for low energies (lower than 2.3 keV, in the case of the detector characterized). In this work, the parameters related to peak asymmetry were studied, allowing a better understanding of the trapping process mentioned above. For this purpose, spectra from mono- and multi-element samples were collected for elements with atomic number between 7 and 20. In order to describe the shape of the characteristic K peaks as a function of its energy, an asymmetric correction to a Gaussian function was proposed. Spectra were obtained by electron probe microanalysis for incidence energies between 5 and 25 keV using an energy dispersive spectrometer equipped with an ultra-thin window Si(Li) detector. It was observed that the area corresponding to the asymmetric correction exhibits an energy dependence similar to that of the mass absorption coefficient of the detector material. In addition, other two spectrometers were used to investigate the dependence of tailing on the detection system. When two spectrometers with the same kind of detector and different pulse processors were compared, peaks were more asymmetric for lower peaking time values. When two different detectors were used, differences were even more important. © 2007 Elsevier B.V. All rights reserved.
Palabras clave: Peak Asymmetry , Si(Li) Detector , Spectral Deconvolution , Spectral Fitting
Ver el registro completo
 
Archivos asociados
Thumbnail
 
Tamaño: 355.3Kb
Formato: PDF
.
Descargar
Licencia
info:eu-repo/semantics/openAccess Excepto donde se diga explícitamente, este item se publica bajo la siguiente descripción: Creative Commons Attribution-NonCommercial-ShareAlike 2.5 Unported (CC BY-NC-SA 2.5)
Identificadores
URI: http://hdl.handle.net/11336/57678
DOI: https://dx.doi.org/10.1016/j.sab.2007.05.005
URL: https://www.sciencedirect.com/science/article/pii/S0584854707001541
Colecciones
Articulos(CCT - PATAGONIA NORTE)
Articulos de CTRO.CIENTIFICO TECNOL.CONICET - PATAGONIA NORTE
Articulos(CINDECA)
Articulos de CENTRO DE INV EN CS.APLICADAS "DR.JORGE J.RONCO"
Citación
Visñovezky, Claudia Analía; Limandri, Silvina Paola; Canafoglia, Maria Elena; Bonetto, Rita Dominga; Trincavelli, Jorge Carlos; Asymmetry of characteristic X-ray peaks obtained by a Si(Li) detector; Pergamon-Elsevier Science Ltd; Spectrochimica Acta Part B: Atomic Spectroscopy; 62; 5; 5-2007; 492-498
Compartir
Altmétricas
 

Enviar por e-mail
Separar cada destinatario (hasta 5) con punto y coma.
  • Facebook
  • X Conicet Digital
  • Instagram
  • YouTube
  • Sound Cloud
  • LinkedIn

Los contenidos del CONICET están licenciados bajo Creative Commons Reconocimiento 2.5 Argentina License

https://www.conicet.gov.ar/ - CONICET

Inicio

Explorar

  • Autores
  • Disciplinas
  • Comunidades

Estadísticas

Novedades

  • Noticias
  • Boletines

Ayuda

Acerca de

  • CONICET Digital
  • Equipo
  • Red Federal

Contacto

Godoy Cruz 2290 (C1425FQB) CABA – República Argentina – Tel: +5411 4899-5400 repositorio@conicet.gov.ar
TÉRMINOS Y CONDICIONES