Mostrar el registro sencillo del ítem

dc.contributor.author
Regiart, Daniel Matias Gaston  
dc.contributor.author
Fernández Baldo, Martín Alejandro  
dc.contributor.author
Spotorno, Viviana G.  
dc.contributor.author
Bertolino, Franco Adrián  
dc.contributor.author
Raba, Julio  
dc.date.available
2016-05-20T20:26:16Z  
dc.date.issued
2013-03  
dc.identifier.citation
Regiart, Daniel Matias Gaston; Fernández Baldo, Martín Alejandro; Spotorno, Viviana G.; Bertolino, Franco Adrián; Raba, Julio; Ultra sensitive microfluidic immunosensor for determination of clenbuterol in bovine hair samples using electrodeposited gold nanoparticles and magnetic microparticles as bio-affinity platform; Elsevier; Biosensors & Bioelectronics; 41; 3-2013; 211-217  
dc.identifier.issn
0956-5663  
dc.identifier.uri
http://hdl.handle.net/11336/5756  
dc.description.abstract
In this article, we report the first integrated microfluidic immunosensor coupled to a screen-printed carbon electrode (SPCE) applied to determination of clenbuterol (CLB) in bovine hair samples. CLB is a member of the β2-agonist drugs which is used in animal production and is banned in Argentine and the European Union. It represents a potential risk and has to be carefully monitored to avoid the illegal use of high amounts of this compound that could result in human food poisoning. In order to perform the CLB detection, the SPCE was modified by gold nanoparticles (AuNPs) electrodeposition. Quantitative determination of CLB was carried out using a competitive indirect immunoassay, method based on the use of anti-CLB antibodies immobilized on magnetic micro particles. The CLB present in bovine hair samples competes immunologically with alkaline phosphatase (AP) enzyme-labeled CLB conjugate for the anti-CLB specific antibodies. Later, p-aminophenyl phosphate was converted to p-aminophenol by AP, and the electroactive product was quantified on AuNPs/SPCE at +0.1 V. The limit of detection for electrochemical method was 0.008 ng mL−1 and the intra- and inter-assay coefficients of variation were below 6%. This being a veterinary control tool very useful for rapid, sensitive and selective detection of CLB in an “in vitro” technique.  
dc.format
application/pdf  
dc.language.iso
eng  
dc.publisher
Elsevier  
dc.rights
info:eu-repo/semantics/openAccess  
dc.rights.uri
https://creativecommons.org/licenses/by-nc-nd/2.5/ar/  
dc.subject
Microfluidic Immunosensor  
dc.subject
Clenbuterol  
dc.subject
Bovine Hair Samples  
dc.subject
Gold Nanoparticles  
dc.subject
Magnetic Micro Particles  
dc.subject.classification
Química Analítica  
dc.subject.classification
Ciencias Químicas  
dc.subject.classification
CIENCIAS NATURALES Y EXACTAS  
dc.title
Ultra sensitive microfluidic immunosensor for determination of clenbuterol in bovine hair samples using electrodeposited gold nanoparticles and magnetic microparticles as bio-affinity platform  
dc.type
info:eu-repo/semantics/article  
dc.type
info:ar-repo/semantics/artículo  
dc.type
info:eu-repo/semantics/publishedVersion  
dc.date.updated
2016-05-16T20:17:34Z  
dc.journal.volume
41  
dc.journal.pagination
211-217  
dc.journal.pais
Países Bajos  
dc.journal.ciudad
Amsterdam  
dc.description.fil
Fil: Regiart, Daniel Matias Gaston. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico San Luis. Instituto de Química de San Luis; Argentina  
dc.description.fil
Fil: Fernández Baldo, Martín Alejandro. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico San Luis. Instituto de Química de San Luis; Argentina  
dc.description.fil
Fil: Spotorno, Viviana G.. Instituto Nacional de Tecnología Agropecuaria. Centro Nacional de Investigaciones Agropecuarias. Centro de Investigación de Recursos Naturales. Instituto de Recursos Biológicos; Argentina  
dc.description.fil
Fil: Bertolino, Franco Adrián. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico San Luis. Instituto de Química de San Luis; Argentina  
dc.description.fil
Fil: Raba, Julio. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico San Luis. Instituto de Química de San Luis; Argentina  
dc.journal.title
Biosensors & Bioelectronics  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/url/http://www.sciencedirect.com/science/article/pii/S0956566312005441  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/doi/http://dx.doi.org/10.1016/j.bios.2012.08.020  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/doi/10.1016/j.bios.2012.08.020