Artículo
A simple and reproducible capacitive electrode
Fecha de publicación:
03/2016
Editorial:
Elsevier
Revista:
Medical Engineering & Physics
ISSN:
1350-4533
Idioma:
Inglés
Tipo de recurso:
Artículo publicado
Clasificación temática:
Resumen
Capacitive Electrodes (CE) allow the acquisition of biopotentials through a dielectric layer, without the use of electrolytes, just by placing them on skin or clothing, but demands front-ends with ultra-high input impedances. This must be achieved while providing a path for bias currents, calling for ultra-high value resistors and special components and construction techniques. A simple CE that uses bootstrap techniques to avoid ultra-high value components and special materials is proposed. When electrodes are placed on the skin; that is, with coupling capacitances CS of around 100 pF, they present a noise level of 3.3 μVRMS in a 0.5-100 Hz bandwidth, which is appropriate for electrocardiography (ECG) measurements. Construction details of the CE and the complete circuit, including a fast recovery feature, are presented.
Palabras clave:
Active Electrodes
,
Insulating Electrodes
,
Non-Contact Measurements
Archivos asociados
Licencia
Identificadores
Colecciones
Articulos(CCT - LA PLATA)
Articulos de CTRO.CIENTIFICO TECNOL.CONICET - LA PLATA
Articulos de CTRO.CIENTIFICO TECNOL.CONICET - LA PLATA
Citación
Spinelli, Enrique Mario; Guerrero, Federico Nicolás; García, Pablo; Haberman, Marcelo Alejandro; A simple and reproducible capacitive electrode; Elsevier; Medical Engineering & Physics; 38; 3; 3-2016; 286-289