dc.identifier.citation
Berli, Claudio Luis Alberto; Kler, Pablo Alejandro; A quantitative model for lateral flow assays; Springer Heidelberg; Microfluidics and Nanofluidics; 20; 7; 7-2016; 104-112
dc.identifier.issn
1613-4982
dc.identifier.uri
http://hdl.handle.net/11336/38252
dc.description.abstract
A simple mathematical model that quantitatively describes the dynamics of analyte capture in lateral flow assays is presented. The formulation accounts for the capillary-driven flow through the porous membrane, the advective transport of analyte, and the immunoreactions that take place in the detection line. Model predictions match the numerical results obtained by computer simulations of the full advection–diffusion–reaction problem in the operating regime of lateral flow assays. The main system parameters were condensed into two dimensionless numbers, namely the relative fluid velocity and the relative analyte concentration. The system is then completely characterized in the space of these critical numbers. The model is also able to describe the time evolution of analyte binding by using alternative timescalings, which discriminate different experimental conditions. The equations reported are practical tools for the design and optimization lateral flow tests, enabling informed decisions on basic questions such as the appropriate flow rate, sample volume, or assay time. Beyond lateral flow assays, the work offers an improved understanding of the underlying physicochemical processes involved in paper-based microfluidics.
dc.format
application/pdf
dc.language.iso
eng
dc.publisher
Springer Heidelberg
dc.rights
info:eu-repo/semantics/openAccess
dc.rights.uri
https://creativecommons.org/licenses/by-nc-sa/2.5/ar/
dc.subject
Lateral Flow Analysis
dc.subject
Optimal Design
dc.subject
Immunoassays
dc.subject
Numerical Simulations
dc.subject.classification
Otras Ciencias Químicas
dc.subject.classification
Ciencias Químicas
dc.subject.classification
CIENCIAS NATURALES Y EXACTAS
dc.title
A quantitative model for lateral flow assays
dc.type
info:eu-repo/semantics/article
dc.type
info:ar-repo/semantics/artículo
dc.type
info:eu-repo/semantics/publishedVersion
dc.date.updated
2018-03-07T15:53:56Z
dc.journal.volume
20
dc.journal.number
7
dc.journal.pagination
104-112
dc.journal.pais
Alemania
dc.journal.ciudad
Heidelberg
dc.description.fil
Fil: Berli, Claudio Luis Alberto. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Santa Fe. Instituto de Desarrollo Tecnológico para la Industria Química. Universidad Nacional del Litoral. Instituto de Desarrollo Tecnológico para la Industria Química; Argentina
dc.description.fil
Fil: Kler, Pablo Alejandro. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Santa Fe. Centro de Investigaciones en Métodos Computacionales. Universidad Nacional del Litoral. Centro de Investigaciones en Métodos Computacionales; Argentina
dc.journal.title
Microfluidics and Nanofluidics
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/url/http://link.springer.com/article/10.1007/s10404-016-1771-9
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/doi/http://dx.doi.org/10.1007/s10404-016-1771-9
Archivos asociados