dc.identifier.citation
Goicoechea, Hector Casimiro; Martínez Galera, María; Gil García, María D.; Teglia, Carla Mariela; Enhanced high-performance liquid chromatography method for the determination of retinoic acid in plasma: Development, optimization and validation; Elsevier; Journal of Chromatography - A; 1353; 6-2014; 40-48  
dc.identifier.issn
0021-9673  
dc.identifier.uri
http://hdl.handle.net/11336/30900  
dc.description.abstract
When determining endogenous compounds in biological samples, the lack of blank or analyte-free matrix samples involves the use of alternative strategies for calibration and quantitation. This article deals with the development, optimization and validation of a high performance liquid chromatography method for the determination of retinoic acid in plasma, obtaining at the same time information about its isomers, taking into account the basal concentration of these endobiotica. An experimental design was used for the optimization of three variables: mobile phase composition, flow rate and column temperature through a central composite design. Four responses were selected for optimization purposes (area under the peaks, quantity of peaks, analysis time and resolution between the first principal peak and the following one). The optimum conditions resulted in a mobile phase consisting of methanol 83.4% (v/v), acetonitrile 0.6% (v/v) and acid aqueous solution 16.0% (v/v); flow rate of 0.68 mL min−1 and an column temperature of 37.10 °C. Detection was performed at 350 nm by a diode array detector. The method was validated following a holistic approach that included not only the classical parameters related to method performance but also the robustness and the expected proportion of acceptable results lying inside predefined acceptability intervals, i.e., the uncertainty of measurements. The method validation results indicated a high selectivity and good precision characteristics that were studied at four concentration levels, with RSD less than 5.0% for retinoic acid (less than 7.5% for the LOQ concentration level), in intra and inter-assay precision studies. Linearity was proved for a range from 0.00489 to 15.109 ng mL−1 of retinoic acid and the recovery, which was studied at four different fortification levels in phuman plasma samples, varied from 99.5% to 106.5% for retinoic acid. The applicability of the method was demonstrated by determining retinoic acid and obtaining information about its isomers in human and frog plasma samples from different origins.  
dc.format
application/pdf  
dc.language.iso
eng  
dc.publisher
Elsevier  
dc.rights
info:eu-repo/semantics/openAccess  
dc.rights.uri
https://creativecommons.org/licenses/by-nc-nd/2.5/ar/  
dc.subject
Endogenous Compounds  
dc.subject
Validation  
dc.subject
Retinoic Acid  
dc.subject
Isomers  
dc.subject.classification
Otras Ciencias Químicas  
dc.subject.classification
Ciencias Químicas  
dc.subject.classification
CIENCIAS NATURALES Y EXACTAS  
dc.title
Enhanced high-performance liquid chromatography method for the determination of retinoic acid in plasma: Development, optimization and validation  
dc.type
info:eu-repo/semantics/article  
dc.type
info:ar-repo/semantics/artículo  
dc.type
info:eu-repo/semantics/publishedVersion  
dc.date.updated
2017-12-18T15:03:05Z  
dc.journal.volume
1353  
dc.journal.pagination
40-48  
dc.journal.pais
Países Bajos  
dc.journal.ciudad
Ámsterdam  
dc.description.fil
Fil: Teglia, Carla Mariela. Universidad Nacional del Litoral. Facultad de Bioquímica y Ciencias Biológicas. Departamento de Química. Cátedra de Química Analítica; Argentina. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Santa Fe; Argentina  
dc.description.fil
Fil: Gil García, María D.. Universidad de Almería. Área de Química Analítica. Departamento de Química y Física; España  
dc.description.fil
Fil: Martínez Galera, María. Universidad de Almería. Área de Química Analítica. Departamento de Química y Física; España  
dc.description.fil
Fil: Goicoechea, Hector Casimiro. Universidad Nacional del Litoral. Facultad de Bioquímica y Ciencias Biológicas. Departamento de Química. Cátedra de Química Analítica; Argentina. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Santa Fe; Argentina  
dc.journal.title
Journal of Chromatography - A  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/url/http://www.sciencedirect.com/science/article/pii/S0021967314000557  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/doi/http://dx.doi.org/10.1016/j.chroma.2014.01.013