Mostrar el registro sencillo del ítem

dc.contributor.author
Irastorza, Ramiro Miguel  
dc.contributor.author
Hadid, Claudio  
dc.contributor.author
Berjano, Enrique  
dc.date.available
2025-03-27T09:50:09Z  
dc.date.issued
2024-07  
dc.identifier.citation
Irastorza, Ramiro Miguel; Hadid, Claudio; Berjano, Enrique; Effect of dispersive electrode position (anterior vs. posterior) in epicardial radiofrequency ablation of ventricular wall: A computer simulation study; Wiley; International Journal for Numerical Methods in Biomedical Engineering; 40; 8; 7-2024; 1-13  
dc.identifier.issn
2040-7939  
dc.identifier.uri
http://hdl.handle.net/11336/257301  
dc.description.abstract
An epicardial approach is often used in radiofrequency (RF) catheter ablation to ablate ventricular tachycardia when an endocardial approach fails. Our objective was to analyze the effect of the position of the dispersive patch (DP) on lesion size using computer modeling during epicardial approach. We compared the posterior position (patient's back), commonly used in clinical practice, to the anterior position (patient's chest). The model considered ventricular wall thicknesses between 4 and 8 mm, and electrode insertion depths between .3 and .7 mm. RF pulses were simulated with 20 W of power for 30 s duration. Statistically significant differences (P < .001) were found between both DP positions in terms of baseline impedance, RF current (at 15 s) and thermal lesion size. The anterior position involved lower impedance (130.8 ± 4.7 vs. 146.2 ± 4.9 Ω) and a higher current (401.5 ± 5.6 vs. 377.5 ± 5.1 mA). The anterior position created lesion sizes larger than the posterior position: 8.9 ± 0.4 vs. 8.4 ± 0.4 mm in maximum width, 8.6 ± 0.4 vs. 8.1 ± 0.4 mm in surface width, and 4.5 ± 0.4 vs. 4.3 ± 0.4 mm in depth. Our results suggest that: (1) the redirection of the RF currents due to repositioning the PD has little impact on lesion size and only affects baseline impedance, and (2) the differences in lesion size are only 0.5 mm wider and 0.2 mm deeper for the anterior position, which does not seem to have a clinical impact in the context of VT ablation.  
dc.format
application/pdf  
dc.language.iso
eng  
dc.publisher
Wiley  
dc.rights
info:eu-repo/semantics/restrictedAccess  
dc.rights.uri
https://creativecommons.org/licenses/by-nc-sa/2.5/ar/  
dc.subject
COMPUTER MODELLING  
dc.subject
DISPERSIVE ELECTRODE  
dc.subject
EPICARDIAL ABLATION  
dc.subject
VENTRICULAR TACHYCARDIA  
dc.subject.classification
Otras Ingeniería Eléctrica, Ingeniería Electrónica e Ingeniería de la Información  
dc.subject.classification
Ingeniería Eléctrica, Ingeniería Electrónica e Ingeniería de la Información  
dc.subject.classification
INGENIERÍAS Y TECNOLOGÍAS  
dc.title
Effect of dispersive electrode position (anterior vs. posterior) in epicardial radiofrequency ablation of ventricular wall: A computer simulation study  
dc.type
info:eu-repo/semantics/article  
dc.type
info:ar-repo/semantics/artículo  
dc.type
info:eu-repo/semantics/publishedVersion  
dc.date.updated
2025-03-26T19:44:02Z  
dc.identifier.eissn
2040-7947  
dc.journal.volume
40  
dc.journal.number
8  
dc.journal.pagination
1-13  
dc.journal.pais
Estados Unidos  
dc.description.fil
Fil: Irastorza, Ramiro Miguel. Universidad Tecnológica Nacional. Facultad Regional La Plata. Departamento de Ingeniería Mecánica; Argentina. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - La Plata. Instituto de Física de Líquidos y Sistemas Biológicos. Universidad Nacional de La Plata. Facultad de Ciencias Exactas. Instituto de Física de Líquidos y Sistemas Biológicos; Argentina  
dc.description.fil
Fil: Hadid, Claudio. Gobierno de la Ciudad Autonoma de Buenos Aires. Hospital General de Agudos Doctor Cosme Argerich.; Argentina  
dc.description.fil
Fil: Berjano, Enrique. Universidad Politécnica de Valencia; España  
dc.journal.title
International Journal for Numerical Methods in Biomedical Engineering  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/url/https://onlinelibrary.wiley.com/doi/10.1002/cnm.3847  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/doi/http://dx.doi.org/10.1002/cnm.3847