Mostrar el registro sencillo del ítem

dc.contributor.author
Fernandez, Hector  
dc.contributor.author
Macho, Juan M.  
dc.contributor.author
Blasco, Jordi  
dc.contributor.author
San Roman, Luis  
dc.contributor.author
Mailaender, Werner  
dc.contributor.author
Serra, Luis  
dc.contributor.author
Larrabide, Ignacio  
dc.date.available
2024-05-27T15:54:40Z  
dc.date.issued
2015-06  
dc.identifier.citation
Fernandez, Hector; Macho, Juan M.; Blasco, Jordi; San Roman, Luis; Mailaender, Werner; et al.; Computation of the change in length of a braided device when deployed in realistic vessel models; Springer; International Journal of Computer Assisted Radiology and Surgery; 10; 10; 6-2015; 1659-1665  
dc.identifier.issn
1861-6429  
dc.identifier.uri
http://hdl.handle.net/11336/236143  
dc.description.abstract
Purpose An important issue in the deployment of braided stents, such as flow diverters, is the change in length, also known as foreshortening, underwent by the device when is released from the catheter into a blood vessel. The position of the distal end is controlled by the interventionist, but know- ing a priori the position of the proximal end of the device is not trivial. In this work, we assess and validate a novel computer method to predict the length that a braided stent will adopt inside a silicon model of an anatomically accurate vessel. Methods Three-dimensional rotational angiography images of aneurysmatic patients were used to generate surface mod- els of the vessels(3Dmeshes) and then create accurate silicon models from them. A braided stent was deployed into each silicon model to measure its length.Thesamestents deployed on the silicon models were virtually deployed on the 3D meshes using the method being evaluated. Results The method was applied to five stent placements on three different silicon models. The length adopted by the real braided device in the silicon models varies between 15 and 30%from the stent length specified by the manufacturer. The final length predicted by themethodwaswithin the esti- mated error of the measured real stent length.Conclusions The method provides, in a few seconds, the length of a braided stent deployed inside a vessel, showing an accurate estimation of the final length for the cases studied. This technique could provide useful information for plan- ning the intervention and improve endovascular treatment of intracranial aneurysms in the future.  
dc.format
application/pdf  
dc.language.iso
eng  
dc.publisher
Springer  
dc.rights
info:eu-repo/semantics/openAccess  
dc.rights.uri
https://creativecommons.org/licenses/by-nc-sa/2.5/ar/  
dc.subject
Desviadores de Frlujo  
dc.subject
Simulacion  
dc.subject
Tratamiento Endovascular  
dc.subject
Aneurismas Intracraneales  
dc.subject.classification
Ciencias de la Información y Bioinformática  
dc.subject.classification
Ciencias de la Computación e Información  
dc.subject.classification
CIENCIAS NATURALES Y EXACTAS  
dc.title
Computation of the change in length of a braided device when deployed in realistic vessel models  
dc.type
info:eu-repo/semantics/article  
dc.type
info:ar-repo/semantics/artículo  
dc.type
info:eu-repo/semantics/publishedVersion  
dc.date.updated
2016-11-23T19:57:17Z  
dc.journal.volume
10  
dc.journal.number
10  
dc.journal.pagination
1659-1665  
dc.journal.pais
Alemania  
dc.journal.ciudad
Berlín  
dc.description.fil
Fil: Fernandez, Hector. Galgo Medical S.L.; España  
dc.description.fil
Fil: Macho, Juan M.. Hospital Clinic Provincial de Barcelona; España  
dc.description.fil
Fil: Blasco, Jordi. Hospital Clinic Provincial de Barcelona; España  
dc.description.fil
Fil: San Roman, Luis. Hospital Clinic Provincial de Barcelona; España  
dc.description.fil
Fil: Mailaender, Werner. Acandis GmbH; Alemania  
dc.description.fil
Fil: Serra, Luis. Galgo Medical S.L.; España  
dc.description.fil
Fil: Larrabide, Ignacio. Universidad Nacional del Centro de la Provincia de Buenos Aires. Facultad de Ciencias Exactas. Grupo de Plasmas Densos Magnetizados. Provincia de Buenos Aires. Gobernación. Comisión de Investigaciones Científicas. Grupo de Plasmas Densos Magnetizados; Argentina. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Tandil; Argentina  
dc.journal.title
International Journal of Computer Assisted Radiology and Surgery  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/url/https://link.springer.com/article/10.1007/s11548-015-1230-1  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/doi/https://doi.org/10.1007/s11548-015-1230-1