Mostrar el registro sencillo del ítem
dc.contributor.author
Fernandez, Hector
dc.contributor.author
Macho, Juan M.
dc.contributor.author
Blasco, Jordi
dc.contributor.author
San Roman, Luis
dc.contributor.author
Mailaender, Werner
dc.contributor.author
Serra, Luis
dc.contributor.author
Larrabide, Ignacio
dc.date.available
2024-05-27T15:54:40Z
dc.date.issued
2015-06
dc.identifier.citation
Fernandez, Hector; Macho, Juan M.; Blasco, Jordi; San Roman, Luis; Mailaender, Werner; et al.; Computation of the change in length of a braided device when deployed in realistic vessel models; Springer; International Journal of Computer Assisted Radiology and Surgery; 10; 10; 6-2015; 1659-1665
dc.identifier.issn
1861-6429
dc.identifier.uri
http://hdl.handle.net/11336/236143
dc.description.abstract
Purpose An important issue in the deployment of braided stents, such as flow diverters, is the change in length, also known as foreshortening, underwent by the device when is released from the catheter into a blood vessel. The position of the distal end is controlled by the interventionist, but know- ing a priori the position of the proximal end of the device is not trivial. In this work, we assess and validate a novel computer method to predict the length that a braided stent will adopt inside a silicon model of an anatomically accurate vessel. Methods Three-dimensional rotational angiography images of aneurysmatic patients were used to generate surface mod- els of the vessels(3Dmeshes) and then create accurate silicon models from them. A braided stent was deployed into each silicon model to measure its length.Thesamestents deployed on the silicon models were virtually deployed on the 3D meshes using the method being evaluated. Results The method was applied to five stent placements on three different silicon models. The length adopted by the real braided device in the silicon models varies between 15 and 30%from the stent length specified by the manufacturer. The final length predicted by themethodwaswithin the esti- mated error of the measured real stent length.Conclusions The method provides, in a few seconds, the length of a braided stent deployed inside a vessel, showing an accurate estimation of the final length for the cases studied. This technique could provide useful information for plan- ning the intervention and improve endovascular treatment of intracranial aneurysms in the future.
dc.format
application/pdf
dc.language.iso
eng
dc.publisher
Springer
dc.rights
info:eu-repo/semantics/openAccess
dc.rights.uri
https://creativecommons.org/licenses/by-nc-sa/2.5/ar/
dc.subject
Desviadores de Frlujo
dc.subject
Simulacion
dc.subject
Tratamiento Endovascular
dc.subject
Aneurismas Intracraneales
dc.subject.classification
Ciencias de la Información y Bioinformática
dc.subject.classification
Ciencias de la Computación e Información
dc.subject.classification
CIENCIAS NATURALES Y EXACTAS
dc.title
Computation of the change in length of a braided device when deployed in realistic vessel models
dc.type
info:eu-repo/semantics/article
dc.type
info:ar-repo/semantics/artículo
dc.type
info:eu-repo/semantics/publishedVersion
dc.date.updated
2016-11-23T19:57:17Z
dc.journal.volume
10
dc.journal.number
10
dc.journal.pagination
1659-1665
dc.journal.pais
Alemania
dc.journal.ciudad
Berlín
dc.description.fil
Fil: Fernandez, Hector. Galgo Medical S.L.; España
dc.description.fil
Fil: Macho, Juan M.. Hospital Clinic Provincial de Barcelona; España
dc.description.fil
Fil: Blasco, Jordi. Hospital Clinic Provincial de Barcelona; España
dc.description.fil
Fil: San Roman, Luis. Hospital Clinic Provincial de Barcelona; España
dc.description.fil
Fil: Mailaender, Werner. Acandis GmbH; Alemania
dc.description.fil
Fil: Serra, Luis. Galgo Medical S.L.; España
dc.description.fil
Fil: Larrabide, Ignacio. Universidad Nacional del Centro de la Provincia de Buenos Aires. Facultad de Ciencias Exactas. Grupo de Plasmas Densos Magnetizados. Provincia de Buenos Aires. Gobernación. Comisión de Investigaciones Científicas. Grupo de Plasmas Densos Magnetizados; Argentina. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Tandil; Argentina
dc.journal.title
International Journal of Computer Assisted Radiology and Surgery
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/url/https://link.springer.com/article/10.1007/s11548-015-1230-1
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/doi/https://doi.org/10.1007/s11548-015-1230-1
Archivos asociados