Mostrar el registro sencillo del ítem

dc.contributor.author
Pollano, Antonella  
dc.contributor.author
Suarez, Marta Magdalena  
dc.contributor.author
Rivarola, María Angélica  
dc.contributor.author
Dominguez, Alejandro  
dc.date.available
2017-08-24T18:22:20Z  
dc.date.issued
2013-02  
dc.identifier.citation
Pollano, Antonella; Suarez, Marta Magdalena; Rivarola, María Angélica; Dominguez, Alejandro; Optimización y Evaluación de un Dispositivo Recuperador de Trombos Arteriales; Institute of Electrical and Electronics Engineers; IEEE Latin America Transactions; 11; 1; 2-2013; 282-287  
dc.identifier.issn
1548-0992  
dc.identifier.uri
http://hdl.handle.net/11336/22937  
dc.description.abstract
The aim of this work was the optimization of an arterial clot retrieval device made of Ni-Ti in the shape of a helical spring. Wire was obteined from GAC International Inc., a company dedicated to the development of products vinculated to orthodontic and orthopedic, and then employed for its development. Finite Element (FE) simulation was used to compare the effects in the clot when different shapes of springs, cylindrical and conical, are used. Abaqus 6.9 was the software chosen to develop those simulations. The assessment of the device using an in vivo model was attempted; an experimental protocol was developed, using the abdominal aorta of the rat for effective in vivo testing of the endovascular catheter-delivered device. Taking into account the results obtained with the FE analysis it was concluded that the utilization of conical shaped springs is better than cilindrical due to the lower stress caused in the clot in this case, that avoids its rupture. The in vivo intervention was succesfull, the generation of the clot in the aorthic artery of a rat was achieved and so was the extraction of it.  
dc.format
application/pdf  
dc.language.iso
spa  
dc.publisher
Institute of Electrical and Electronics Engineers  
dc.rights
info:eu-repo/semantics/openAccess  
dc.rights.uri
https://creativecommons.org/licenses/by-nc-sa/2.5/ar/  
dc.subject
Mechanical Embolectomy  
dc.subject
Nitinol  
dc.subject
Finite Element Method.  
dc.subject.classification
Bioquímica y Biología Molecular  
dc.subject.classification
Medicina Básica  
dc.subject.classification
CIENCIAS MÉDICAS Y DE LA SALUD  
dc.title
Optimización y Evaluación de un Dispositivo Recuperador de Trombos Arteriales  
dc.type
info:eu-repo/semantics/article  
dc.type
info:ar-repo/semantics/artículo  
dc.type
info:eu-repo/semantics/publishedVersion  
dc.date.updated
2017-08-22T18:11:28Z  
dc.journal.volume
11  
dc.journal.number
1  
dc.journal.pagination
282-287  
dc.journal.pais
Estados Unidos  
dc.description.fil
Fil: Pollano, Antonella. Universidad Nacional de Córdoba; Argentina. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina  
dc.description.fil
Fil: Suarez, Marta Magdalena. Universidad Nacional de Córdoba; Argentina. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina  
dc.description.fil
Fil: Rivarola, María Angélica. Universidad Nacional de Córdoba; Argentina. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina  
dc.description.fil
Fil: Dominguez, Alejandro. Universidad Nacional de Córdoba; Argentina. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina  
dc.journal.title
IEEE Latin America Transactions  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/url/http://www.ewh.ieee.org/reg/9/etrans/ieee/issues/vol11/vol11issue1Feb.2013/49Pollano.htm