Córdoba, Luis Pedro; Gomez Lopez, Maria de Los Angeles; Goy, Carla Belen; Herrera de Olivera, Myriam Cristina; Design and Evaluation of Pressure and Electrocardiographic Modules Embedded in a Cardiac Hemodynamic Monitor ; Institute of Electrical and Electronics Engineers; IEEE Latin America Transactions; 11; 1; 2-2013; 107-113  
dc.identifier.issn
1548-0992  
dc.identifier.uri
http://hdl.handle.net/11336/22012  
dc.description.abstract
This work presents the design and the implementation of a two-channels recording system and its adaptation to a FPGA-based acquisition system. The recorded signals are Electrocardiogram (ECG) and Intraventricular Pressure (IVP). Elements and circuits needed to incorporate the newly developed recording system to a previously developed FPGA-based Conductance Catheter System (SCC) are also presented. The result is a fully integrated system, in which the evaluation of each channel is carried out. ECG amplifier is evaluated by means of frequency response and common mode rejection ratio. Filter frequency response and two calibration methods, over 0-150 mmHg and 150-300 mmHg ranges, are used to evaluate PIV channel. Finally analogical to digital conversion and communication between two-channels recording system and FPGA-based acquisition system are tested. Satisfactory results are obtained.  
dc.format
application/pdf  
dc.language.iso
spa  
dc.publisher
Institute of Electrical and Electronics Engineers  
dc.rights
info:eu-repo/semantics/openAccess  
dc.rights.uri
https://creativecommons.org/licenses/by-nc-sa/2.5/ar/  
dc.subject
Intraventricular Pressure  
dc.subject
Ecg  
dc.subject
Fpga  
dc.subject
Conductance Catheter System  
dc.subject.classification
Ingeniería Médica  
dc.subject.classification
Ingeniería Médica  
dc.subject.classification
INGENIERÍAS Y TECNOLOGÍAS  
dc.title
Design and Evaluation of Pressure and Electrocardiographic Modules Embedded in a Cardiac Hemodynamic Monitor  
dc.type
info:eu-repo/semantics/article  
dc.type
info:ar-repo/semantics/artículo  
dc.type
info:eu-repo/semantics/publishedVersion  
dc.date.updated
2017-08-07T17:37:24Z  
dc.journal.volume
11  
dc.journal.number
1  
dc.journal.pagination
107-113  
dc.journal.pais
Estados Unidos  
dc.journal.ciudad
Nueva York  
dc.description.fil
Fil: Córdoba, Luis Pedro.  
dc.description.fil
Fil: Gomez Lopez, Maria de Los Angeles. Universidad Nacional de Tucuman. Facultad de Ciencias Exactas y Tecnología. Departamento de Electricidad, Electrónica y Computación; Argentina  
dc.description.fil
Fil: Goy, Carla Belen. Universidad Nacional de Tucuman. Facultad de Ciencias Exactas y Tecnología. Departamento de Electricidad, Electrónica y Computación; Argentina. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina  
dc.description.fil
Fil: Herrera de Olivera, Myriam Cristina. Universidad Nacional de Tucumán. Facultad de Ciencias Exactas y Tecnologia; Argentina. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina  
dc.journal.title
IEEE Latin America Transactions  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/doi/http://dx.doi.org/10.1109/TLA.2013.6502786  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/url/http://ieeexplore.ieee.org/document/6502786/