Mostrar el registro sencillo del ítem

dc.contributor.author
Conde Garrido, Juan Manuel  
dc.contributor.author
Couselo, M. A.  
dc.contributor.author
Silveyra, Josefina María  
dc.date.available
2023-12-04T14:43:24Z  
dc.date.issued
2023-01  
dc.identifier.citation
Conde Garrido, Juan Manuel; Couselo, M. A.; Silveyra, Josefina María; Primer prototipo de trampa fría con pasante eléctrico para vacío de bajo costo; Asociación Física Argentina; Anales AFA; 34; 1; 1-2023; 8-16  
dc.identifier.issn
0327-358X  
dc.identifier.uri
http://hdl.handle.net/11336/219196  
dc.description.abstract
En un trabajo previo, Conde Garrido y Silveyra propusieron una novedosa tecnología de trampa fría capaz de atrapar contaminantes en sistemas de vacío. La misma fue diseñada para ser aplicada en sistemas de deposición de películas delgadas de vidrios calcogenuros por ablación láser pulsada. Mientras que los sistemas tradicionales se enfrían por compresión de gases o baños fríos como nitrógeno líquido, la trampa reportada se enfría por efecto termoeléctrico. Esto permite reducir la inversión de capital, los costos operativos y los tiempos de puesta en marcha y mantenimiento. En el presente trabajo se presenta la construcción, puesta a punto y caracterización del primer prototipo físico de la trampa fría. La caracterización incluyó, en primer lugar, el control de la presión final alcanzada por el dispositivo. Para caracterizar el desempeño térmico del dispositivo, se diseñó y fabricó un sistema de medición temperaturas. Como parte de este sistema de medición, se destaca un pasante eléctrico para vacío de bajo costo. Los resultados indican que, con la trampa fabricada, se pueden alcanzar presiones menores a 2 × 10−5 mbar, mientras que las superficies frías alcanzan temperaturas de aproximadamente −12 ◦C. Los niveles de vacío y de temperatura cumplen las condiciones requeridas para el sistema de deposición de películas delgadas de vidrios calcogenuros por ablación láser pulsada. Sin embargo, no se logran temperaturas tan bajas como las estimadas para el prototipo virtual (de hasta −50◦C). Se identificaron entonces los puentes y resistencias térmicas presentes en el dispositivo fabricado, despreciados en el modelo, señalando oportunidades de mejora. Finalmente, se proponen revisiones al diseño actual que simplifican su proceso de manufactura, mejoran su robustez y eficacia, y facilitan su operación y mantenimiento.  
dc.description.abstract
In a previous work, Conde Garrido and Silveyra proposed a novel cold trap (baffle) technology capable of trapping contaminants in vacuum systems. The baffle was designed to be applied in systems to synthesize chalcogenide glass thin films by pulsed laser deposition. While traditional baffles are cooled down with compression cooling systems or cooling solutions such as liquid nitrogen, the reported baffle is cooled down by the thermoelectric effect, which allows for reducing the capital investment, operating costs, as well as start-up and maintenance times. This paper presents the construction, tuning, and characterization of the first physical prototype of the baffle. The characterization included, first, the control of the final pressure reached by the device. To characterize the thermal performance of the baffle, a temperature measurement system was designed and manufactured. Within this measurement system, we highlight a low-cost electric vacuum feedthrough. The results indicate that the constructed baffle can reach pressures lower than 2 × 10−5 mbar, while the cold surfaces reach temperatures of approximately −12◦ C. The vacuum and cold temperature levels meet the required conditions for the pulsed laser deposition of chalcogenide glass thin films. However, temperatures are not as low as those estimated for the virtual prototype (down to −50◦ C). Thermal bridges and resistances present in the fabricated device, neglected in the model, were then identified, pointing out opportunities for improvement. Finally, revisions to the current design are proposed that simplify its manufacturing process, improve its robustness and efficiency, and facilitate its operation and maintenance.  
dc.format
application/pdf  
dc.language.iso
spa  
dc.publisher
Asociación Física Argentina  
dc.rights
info:eu-repo/semantics/openAccess  
dc.rights.uri
https://creativecommons.org/licenses/by-nc/2.5/ar/  
dc.subject
BAFFLE  
dc.subject
THERMOELECTRIC  
dc.subject
ELECTRICAL FEEDTHROUGH  
dc.subject
VACUUM  
dc.subject.classification
Nano-materiales  
dc.subject.classification
Nanotecnología  
dc.subject.classification
INGENIERÍAS Y TECNOLOGÍAS  
dc.title
Primer prototipo de trampa fría con pasante eléctrico para vacío de bajo costo  
dc.title
First prototype of a low-cost vacuum baffle with electrical feedthrough  
dc.type
info:eu-repo/semantics/article  
dc.type
info:ar-repo/semantics/artículo  
dc.type
info:eu-repo/semantics/publishedVersion  
dc.date.updated
2023-11-29T13:22:47Z  
dc.identifier.eissn
1850-1168  
dc.journal.volume
34  
dc.journal.number
1  
dc.journal.pagination
8-16  
dc.journal.pais
Argentina  
dc.journal.ciudad
Tandil  
dc.description.fil
Fil: Conde Garrido, Juan Manuel. Consejo Nacional de Investigaciones Científicas y Técnicas. Oficina de Coordinación Administrativa Houssay. Instituto de Tecnologías y Ciencias de la Ingeniería "Hilario Fernández Long". Universidad de Buenos Aires. Facultad de Ingeniería. Instituto de Tecnologías y Ciencias de la Ingeniería "Hilario Fernández Long"; Argentina. Universidad de Buenos Aires. Facultad de Ingeniería. Departamento de Física. Laboratorio de Sólidos Amorfos; Argentina  
dc.description.fil
Fil: Couselo, M. A.. Universidad de Buenos Aires. Facultad de Ingeniería. Departamento de Física. Laboratorio de Sólidos Amorfos; Argentina  
dc.description.fil
Fil: Silveyra, Josefina María. Consejo Nacional de Investigaciones Científicas y Técnicas. Oficina de Coordinación Administrativa Houssay. Instituto de Tecnologías y Ciencias de la Ingeniería "Hilario Fernández Long". Universidad de Buenos Aires. Facultad de Ingeniería. Instituto de Tecnologías y Ciencias de la Ingeniería "Hilario Fernández Long"; Argentina. Universidad de Buenos Aires. Facultad de Ingeniería. Departamento de Física. Laboratorio de Sólidos Amorfos; Argentina  
dc.journal.title
Anales AFA  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/url/https://anales.fisica.org.ar/index.php/analesafa/article/view/2365  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/doi/https://doi.org/10.31527/analesafa.2023.34.1.8