Mostrar el registro sencillo del ítem
dc.contributor.author
Rodríguez Sarmiento, David L.
dc.contributor.author
León Vargas, Fabian
dc.contributor.author
Garelli, Fabricio
dc.date.available
2023-07-28T12:48:59Z
dc.date.issued
2022-06
dc.identifier.citation
Rodríguez Sarmiento, David L.; León Vargas, Fabian; Garelli, Fabricio; Practical constraint definition in safety schemes for artificial pancreas systems; Wichtig Editore; International Journal Of Artificial Organs; 45; 6; 6-2022; 535-542
dc.identifier.issn
0391-3988
dc.identifier.uri
http://hdl.handle.net/11336/205916
dc.description.abstract
Introduction: Artificial pancreas systems usually define an insulin-on-board constraint ((Formula presented.)) for safety schemes to limit the insulin infusion and avoid hypoglycemia during the closed-loop performance. Several methods have been proposed with impractical considerations requiring information from the prandial events or complex procedures for ambulatory use. Methods: This paper presents a simple method that consists of two novel rules that allow finding an (Formula presented.) based only on common clinical parameters that do not require patient intervention. The method robustness was evaluated using a control system coupled to a safety layer under demanding scenarios implemented on the FDA-approved simulator for preclinical studies. Results: The method maintains a safe performance, even in the face of interpatient variability, hybrid and fully automatic implementations of an artificial pancreas system, and uncertain settings. Both proposed rules work as effectively or even better and without the patient intervention than other methods that have already been clinically validated. Conclusion: This method can be used to define a constant (Formula presented.) that ensures performance and safety of the control system, even under scenarios with incorrect clinical data. Unlike other methods, this method only requires reliable information that is easily obtained from the patient, such as their total daily dose of insulin or body mass.
dc.format
application/pdf
dc.language.iso
eng
dc.publisher
Wichtig Editore
dc.rights
info:eu-repo/semantics/restrictedAccess
dc.rights.uri
https://creativecommons.org/licenses/by-nc-sa/2.5/ar/
dc.subject
ARTIFICIAL PANCREAS
dc.subject
HYPOGLYCEMIA
dc.subject
IN-SILICO ASSAY
dc.subject
INSULIN ON BOARD
dc.subject
TYPE 1 DIABETES
dc.subject.classification
Sistemas de Automatización y Control
dc.subject.classification
Ingeniería Eléctrica, Ingeniería Electrónica e Ingeniería de la Información
dc.subject.classification
INGENIERÍAS Y TECNOLOGÍAS
dc.title
Practical constraint definition in safety schemes for artificial pancreas systems
dc.type
info:eu-repo/semantics/article
dc.type
info:ar-repo/semantics/artículo
dc.type
info:eu-repo/semantics/publishedVersion
dc.date.updated
2023-07-07T18:02:40Z
dc.journal.volume
45
dc.journal.number
6
dc.journal.pagination
535-542
dc.journal.pais
Italia
dc.journal.ciudad
Milán (Italia)
dc.description.fil
Fil: Rodríguez Sarmiento, David L.. Universidad Antonio Nariño; Colombia
dc.description.fil
Fil: León Vargas, Fabian. Universidad Antonio Nariño; Colombia
dc.description.fil
Fil: Garelli, Fabricio. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - La Plata. Instituto de Investigaciones en Electrónica, Control y Procesamiento de Señales. Universidad Nacional de La Plata. Instituto de Investigaciones en Electrónica, Control y Procesamiento de Señales; Argentina
dc.journal.title
International Journal Of Artificial Organs
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/doi/http://dx.doi.org/10.1177/03913988221095586
Archivos asociados