Evaluation of an automated chemiluminescent immunoassay for salivary cortisol measurement. Utility in the diagnosis of Cushing's syndrome
Repetto, Esteban Martín
; Gonzalez, Diego Javier; Jacobsen, Dario; Smithuis, Fernando; Jamardo, Juan; Cano, Mariel; Aranda, Claudio; Oneto, Adriana; Berg, Gabriela Alicia
; Fabre, Bibiana
Fecha de publicación:
03/2017
Editorial:
De Gruyter
Revista:
Clinical Chemistry and Laboratory Medicine
ISSN:
1434-6621
Idioma:
Inglés
Tipo de recurso:
Artículo publicado
Clasificación temática:
Resumen
The aim of the current study was to validate an automated chemiluminescent immunoassay for cortisol quantification in saliva and to determine the cut-off level for Cushing´s Syndrome diagnosis in an adult population. The salivary cortisol levels from healthy individuals and patients previously diagnosed with Cushing´s Syndrome were measured with Siemens Immulite 2000 ® automatic chemiluminescent enzyme-immunoassay analyser with a brief modification. The limit of quantification was 3,4 nmol/L. The assay also demonstrated a high level of accuracy, as determined by linearity under dilution and recovery tests. Using this methodology, levels higher than 5.7 nmol/L is indicative of Cushing´s Syndrome with a diagnostic sensitivity and specificity of 90,5% and 97.4% respectively. The assay validated in the present study could be used in the clinical laboratories for the evaluation of cortisol levels in patients in which Cushing´s Syndrome is suspected.
Palabras clave:
Chemiluminescence
,
Cortisol
,
Cushing'S Syndrome
,
Saliva
Archivos asociados
Licencia
Identificadores
Colecciones
Articulos(CEFYBO)
Articulos de CENTRO DE ESTUDIOS FARMACOLOGICOS Y BOTANICOS
Articulos de CENTRO DE ESTUDIOS FARMACOLOGICOS Y BOTANICOS
Citación
Repetto, Esteban Martín; Gonzalez, Diego Javier; Jacobsen, Dario; Smithuis, Fernando; Jamardo, Juan; et al.; Evaluation of an automated chemiluminescent immunoassay for salivary cortisol measurement. Utility in the diagnosis of Cushing's syndrome; De Gruyter; Clinical Chemistry and Laboratory Medicine; 55; 3; 3-2017; 65-68
Compartir
Altmétricas