Mostrar el registro sencillo del ítem

dc.contributor.author
Barreiro, Martina  
dc.contributor.author
Sánchez, Pablo  
dc.contributor.author
Vera, Julián  
dc.contributor.author
Viera, Matías  
dc.contributor.author
Morales, Isabel  
dc.contributor.author
Dell'osa, Antonio Héctor  
dc.contributor.author
Bertemes Filho, Pedro  
dc.contributor.author
Simini, Franco  
dc.date.available
2023-04-17T16:12:18Z  
dc.date.issued
2022-03  
dc.identifier.citation
Barreiro, Martina; Sánchez, Pablo; Vera, Julián; Viera, Matías; Morales, Isabel; et al.; Multiplexing Error and Noise Reduction in Electrical Impedance Tomography Imaging; Frontiers Media; Frontiers in Electronics; 3; 1; 3-2022; 1-11  
dc.identifier.uri
http://hdl.handle.net/11336/194203  
dc.description.abstract
Electrical Impedance Tomography design can be simplified to obtain a low cost 16 electrodes edema monitoring clinical instrument by using voltage measurement multiplexing. Multiplexers introduce errors, which we have estimated by consecutive phantom measurements both using voltage multiplexers and by selecting the electrodes by hand, all other things being the same. Noise is taken care of by averaging. The EIDORS reconstruction of the phantom with multiplexed measurements is compared to the hand-selected electrode measurements reconstruction. The difference image obtained is considered an estimation of the multiplexer induced error. This measurement error is subtracted from the multiplexed object measurement matrix, giving a modified reconstruction which is closer to the hand-selected electrodes measurement based reconstruction than the multiplexed reconstruction. The quality factor of the uncorrected multiplexer obtained image of 57% is increased to 83% which is the best increase of three methods described. This suggests the benefit of a “calibration” phase for all 16 electrodes, prior to EIT reconstruction, using a set-up-specific “error matrix” to correct the data matrix before submission to the reconstruction method.  
dc.format
application/pdf  
dc.language.iso
eng  
dc.publisher
Frontiers Media  
dc.rights
info:eu-repo/semantics/restrictedAccess  
dc.rights.uri
https://creativecommons.org/licenses/by-nc-sa/2.5/ar/  
dc.subject
EIT  
dc.subject
PHANTOM  
dc.subject
MULTIPLEXER  
dc.subject
NOISE REDUCTION  
dc.subject
EIDORS  
dc.subject
ERROR REDUCTION  
dc.subject.classification
Otras Ingeniería Médica  
dc.subject.classification
Ingeniería Médica  
dc.subject.classification
INGENIERÍAS Y TECNOLOGÍAS  
dc.title
Multiplexing Error and Noise Reduction in Electrical Impedance Tomography Imaging  
dc.type
info:eu-repo/semantics/article  
dc.type
info:ar-repo/semantics/artículo  
dc.type
info:eu-repo/semantics/publishedVersion  
dc.date.updated
2023-04-17T11:08:58Z  
dc.identifier.eissn
2673-5857  
dc.journal.volume
3  
dc.journal.number
1  
dc.journal.pagination
1-11  
dc.journal.pais
Suiza  
dc.journal.ciudad
Lausana  
dc.description.fil
Fil: Barreiro, Martina. Universidad de la Republica. Facultad de Ingeniería; Uruguay  
dc.description.fil
Fil: Sánchez, Pablo. Universidad de la Republica. Facultad de Ingeniería; Uruguay  
dc.description.fil
Fil: Vera, Julián. Universidad de la Republica. Facultad de Ingeniería; Uruguay  
dc.description.fil
Fil: Viera, Matías. Universidad de la Republica. Facultad de Ingeniería; Uruguay  
dc.description.fil
Fil: Morales, Isabel. Universidad de la Republica. Facultad de Ingeniería; Uruguay  
dc.description.fil
Fil: Dell'osa, Antonio Héctor. Universidad Nacional de Tierra del Fuego. Instituto de Desarrollo Economico E Innovacion; Argentina. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina  
dc.description.fil
Fil: Bertemes Filho, Pedro. Universidade Federal de Santa Catarina; Brasil  
dc.description.fil
Fil: Simini, Franco. Universidad de la Republica. Facultad de Ingeniería; Uruguay  
dc.journal.title
Frontiers in Electronics  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/url/https://www.frontiersin.org/articles/10.3389/felec.2022.848618/full  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/doi/http://dx.doi.org/10.3389/felec.2022.848618