Artículo
Experimental artifact in MOKE measurements when using paramagnetic sample holders
Fecha de publicación:
07/2020
Editorial:
IOP Publishing
Revista:
Measurement Science & Technology (print)
ISSN:
0957-0233
Idioma:
Inglés
Tipo de recurso:
Artículo publicado
Clasificación temática:
Resumen
We describe here an artifact that may affect to magneto-optical Kerr measurements. When paramagnetic sample holders (SH) with non-negligible susceptibilities are used, the inhomogeneity of the applied magnetic field can induce forces and torques on it, shifting the reflected beam, and altering its intensity at the photodetector. The effect is even and can be avoided using low susceptibility paramagnetic or diamagnetic SH We also present a detailed analytical description of the magnetic forces involved and provide some estimated values of the SH shifting, showing that they might distort the magneto-optical Kerr effect signal. Moreover, in this paper we show how the artifact can be removed from the experimental curves with an appropriated data analysis.
Palabras clave:
DATA ANALYSIS
,
EXPERIMENTAL ARTIFACTS
,
MAGNETO-OPTIC KERR EFFECT
Archivos asociados
Licencia
Identificadores
Colecciones
Articulos(IFIS - LITORAL)
Articulos de INST.DE FISICA DEL LITORAL
Articulos de INST.DE FISICA DEL LITORAL
Citación
Munoz-Noval, Alvaro; Bonin, Claudio Julio; Bonetto, Fernando Jose; Garcia, Miguel Angel; Experimental artifact in MOKE measurements when using paramagnetic sample holders; IOP Publishing; Measurement Science & Technology (print); 31; 12; 7-2020
Compartir
Altmétricas