Mostrar el registro sencillo del ítem

dc.contributor.author
Oliva, Matias Javier  
dc.contributor.author
Garcia, Pablo Andres  
dc.contributor.author
Spinelli, Enrique Mario  
dc.date.available
2021-09-17T19:32:49Z  
dc.date.issued
2020  
dc.identifier.citation
Oliva, Matias Javier; Garcia, Pablo Andres; Spinelli, Enrique Mario; A system on chip based electroencephalogram acquisition system; Sociedad Argentina de Bioingeniería; Revista Argentina de Bioingeniería; 24; 3; 2020; 8-12  
dc.identifier.issn
0329-5257  
dc.identifier.uri
http://hdl.handle.net/11336/140771  
dc.description.abstract
La mayoría de las implementaciones actuales de interfaces cerebro computadora (BCI) consisten en una etapa de adquisición de biopotenciales y una PC, donde se realiza el procesamiento de las señales. Como estas han demostrado su potencialidad para mejorar la calidad de vida de personas con movilidad reducida y pérdida del habla se plantea la necesidad de migrar a un dispositivo de tamaño reducido, eliminando la PC del esquema. Los estrictos requerimientos de procesamiento en tiempo real de las BCI justifican la elección de un sistema embebido heterogéneo para este propósito. En este trabajo se presenta un adquisidor de señales de electroencefalograma (EEG) basado en un sistema SoC (System on Chip) DE10-nano de bajo costo, provisto por Altera. Se describe el sistema por completo y se muestran registros de EEG de un ritmo alfa, correspondientes a un usuario durante un periodo de relajación. Mediante el procesamiento off line de la señal se obtiene su espectro de frecuencias, el cual tiene un máximo de amplitud en 12,8 Hz, característico de una señal de este tipo. La validación del sistema de adquisición implementado constituye un punto de partida sólido para el desarrollo de una BCI completamente funcional en el futuro, así como otro tipo de aplicaciones con restricciones temporales más demandantes.  
dc.description.abstract
Current Brain computer interfaces (BCI) are usually implemented by a biopotential acquisition board and a computer (PC), where the signal processing is performed. Since these have demonstrated their potential to improve the quality of life of people with reduced mobility and speech losses, it is time to migrate to smaller devices, eliminating the PC from the scheme. The strict real time processing requirements of BCI justify the use of a System on Chip (SoC) for this propose. This work presents an electroencephalogram (EEG) acquisition system based on a low-cost DE10-nano SoC provided by Altera. The complete system is described and EEG records of an alpha rhythm, corresponding to a user during a period of relaxation, are shown. Through offline processing of the signal, its frequency spectrum is obtained, which has a maximum amplitude at 12.8 Hz, characteristic for a signal of this type. The validation of the acquisition system constitutes a solid starting point for the development of a fully functional BCI in the future, as well as other applications with more demanding time constraints.an electroencephalogram (EEG) acquisition system based on a low-cost DE10-nano SoC provided by Altera. The complete system is described and EEG records of an alpha rhythm, corresponding to a user during a period of relaxation, are shown. Through offline processing of the signal, its frequency spectrum is obtained, which has a maximum amplitude at 12.8 Hz, characteristic for a signal of this type. The validation of the acquisition system constitutes a solid starting point for the development of a fully functional BCI in the future, as well as other applications with more demanding time constraints.  
dc.format
application/pdf  
dc.language.iso
spa  
dc.publisher
Sociedad Argentina de Bioingeniería  
dc.rights
info:eu-repo/semantics/openAccess  
dc.rights.uri
https://creativecommons.org/licenses/by-nc-sa/2.5/ar/  
dc.subject
ELECTROENCEFALOGRAMA  
dc.subject
SISTEMA EMBEBIDO HETEROGENEO  
dc.subject
INTERFAZ CEREBRO COMPUTADORA  
dc.subject.classification
Ingeniería Eléctrica y Electrónica  
dc.subject.classification
Ingeniería Eléctrica, Ingeniería Electrónica e Ingeniería de la Información  
dc.subject.classification
INGENIERÍAS Y TECNOLOGÍAS  
dc.title
A system on chip based electroencephalogram acquisition system  
dc.type
info:eu-repo/semantics/article  
dc.type
info:ar-repo/semantics/artículo  
dc.type
info:eu-repo/semantics/publishedVersion  
dc.date.updated
2021-09-06T17:26:31Z  
dc.identifier.eissn
2591-376X  
dc.journal.volume
24  
dc.journal.number
3  
dc.journal.pagination
8-12  
dc.journal.pais
Argentina  
dc.journal.ciudad
San Miguel de Tucumán  
dc.description.fil
Fil: Oliva, Matias Javier. Universidad Nacional de La Plata. Facultad de Ingeniería; Argentina  
dc.description.fil
Fil: Garcia, Pablo Andres. Universidad Nacional de La Plata. Facultad de Ingeniería; Argentina  
dc.description.fil
Fil: Spinelli, Enrique Mario. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - La Plata. Instituto de Investigaciones en Electrónica, Control y Procesamiento de Señales. Universidad Nacional de La Plata. Instituto de Investigaciones en Electrónica, Control y Procesamiento de Señales; Argentina  
dc.journal.title
Revista Argentina de Bioingeniería  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/url/http://revista.sabi.org.ar/index.php/revista/article/view/353