Mostrar el registro sencillo del ítem

dc.contributor.author
Navarro, Lucila  
dc.contributor.author
Duffó, Gustavo Sergio  
dc.contributor.author
Vetcher, David  
dc.contributor.author
Moles, Victor  
dc.contributor.author
Luna, Julio Alberto  
dc.contributor.author
Rintoul, Ignacio  
dc.date.available
2020-06-08T20:29:03Z  
dc.date.issued
2020-01  
dc.identifier.citation
Navarro, Lucila; Duffó, Gustavo Sergio; Vetcher, David; Moles, Victor; Luna, Julio Alberto; et al.; Coating induced corrosion of coronary stents - A comparative study with clinical consequences; Pharma Project Group; Trends in Medicine; 6; 1-2020; 220-226  
dc.identifier.issn
1594-2848  
dc.identifier.uri
http://hdl.handle.net/11336/106932  
dc.description.abstract
The clinical failsof gold (Au) and carbon (C) coated 316LSS stents have been barelyinvestigated. A comprehensive understanding of these fails maycontribute to the improvement of protocols for the development ofbetter materials and surfaces. This work elucidates the reasons ofthe clinical fails of Au and C coated 316LSS stents, demonstratesthe physical-chemical processes that leaded to these fails, reviewsthe mistaken line of thought that pushed these technologies to themost advanced clinical stages and proposes complementary techniquesto the FDA guidances for the early non-clinical engineering tests. Auand C coated 316LSS, bare-metal 316LSS and Co-Cr stents were studiedusing potentiodynamic scans according to FDA guidance protocols,image analysis and bulk elemental composition. The statisticalclinical outcome of the stents was found to correlate with the ionsrelease patterns eluted from their corresponding metallic substrates.In all cases ions release were consequence of post-passive corrosionprocesses such as pitting corrosion of 316LSS and bulk graindissolution of Co-Cr alloy. Surprisingly, the stents outcome did notcorrelate with their corresponding corrosion rates. Clinical fails ofAu and C coated 316LSS stents are explained in terms of their failsto inhibit the ions release from their 316LSS substrates. Materialsand surface improvements must be oriented to inhibit post-passivecorrosion processes of stents substrates. Potentiodynamic scanscomplemented with image analysis are strongly recommended for theevaluation of pre-clinical engineering tests.  
dc.format
application/pdf  
dc.language.iso
eng  
dc.publisher
Pharma Project Group  
dc.rights
info:eu-repo/semantics/openAccess  
dc.rights.uri
https://creativecommons.org/licenses/by-nc-sa/2.5/ar/  
dc.subject
ALLERGY  
dc.subject
BIOCOMPATIBILITY  
dc.subject
BIOMATERIAL  
dc.subject
CORROSION  
dc.subject
IMPLANT  
dc.subject
STAINLESS STEEL  
dc.subject.classification
Ingeniería Médica  
dc.subject.classification
Ingeniería Médica  
dc.subject.classification
INGENIERÍAS Y TECNOLOGÍAS  
dc.subject.classification
Bioproductos, Biomateriales, Bioplásticos, Biocombustibles, Bioderivados, etc.  
dc.subject.classification
Biotecnología Industrial  
dc.subject.classification
INGENIERÍAS Y TECNOLOGÍAS  
dc.subject.classification
Recubrimientos y Películas  
dc.subject.classification
Ingeniería de los Materiales  
dc.subject.classification
INGENIERÍAS Y TECNOLOGÍAS  
dc.title
Coating induced corrosion of coronary stents - A comparative study with clinical consequences  
dc.type
info:eu-repo/semantics/article  
dc.type
info:ar-repo/semantics/artículo  
dc.type
info:eu-repo/semantics/publishedVersion  
dc.date.updated
2020-06-08T15:37:23Z  
dc.journal.volume
6  
dc.journal.pagination
220-226  
dc.journal.pais
Estados Unidos  
dc.description.fil
Fil: Navarro, Lucila. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Santa Fe. Instituto de Desarrollo Tecnológico para la Industria Química. Universidad Nacional del Litoral. Instituto de Desarrollo Tecnológico para la Industria Química; Argentina  
dc.description.fil
Fil: Duffó, Gustavo Sergio. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina. Comisión Nacional de Energía Atómica; Argentina  
dc.description.fil
Fil: Vetcher, David. Bioimagen; Argentina  
dc.description.fil
Fil: Moles, Victor. Unidad de Intervenciones Cardiovasculares; Argentina  
dc.description.fil
Fil: Luna, Julio Alberto. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Santa Fe. Instituto de Desarrollo Tecnológico para la Industria Química. Universidad Nacional del Litoral. Instituto de Desarrollo Tecnológico para la Industria Química; Argentina  
dc.description.fil
Fil: Rintoul, Ignacio. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Santa Fe. Instituto de Desarrollo Tecnológico para la Industria Química. Universidad Nacional del Litoral. Instituto de Desarrollo Tecnológico para la Industria Química; Argentina  
dc.journal.title
Trends in Medicine  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/url/https://www.oatext.com/coating-induced-corrosion-of-coronary-stents-a-comparative-study-with-clinical-consequences.php  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/doi/http://dx.doi.org/10.15761/TiM.1000220