Artículo
Determination of cobalt in urine by FI-ICP-AES with online preconcentration
Farias, Gustavo Marcelo; Wuilloud, Rodolfo German
; Moyano, Susana; Gasquez, José Antonio; Olsina, Roberto Antonio
; Martinez, Luis Dante
Fecha de publicación:
01/09/2002
Editorial:
Preston Publ Inc
Revista:
Journal Of Analytical Toxicology
ISSN:
0146-4760
e-ISSN:
1945-2403
Idioma:
Inglés
Tipo de recurso:
Artículo publicado
Clasificación temática:
Resumen
A method for the preconcentration and determination of cobalt in human urine samples was developed. The online preconcentration and determination were attained using inductively coupled plasma atomic emission spectromety (ICP-AES) coupled to a flow injection (FI) method. Cobalt was retained on an Amberlite XAD-7 resin as cobalt-2-(5-bromo-2-pyridylazo)-5-diethylaminophenol complex at pH 9.5. Cobalt was removed from the microcolumn with perchloric acid. A sensitivity enhancement factor of 90 was obtained with respect to cobalt determination by ICP-AES without preconcentration. The value of detection limit for the preconcentration method proposed was 25 ng/L. The precision for 10 replicate determinations at the 5 mg/L (mean ± SD, 5.1 ± 0.14) Co level was 2.7% relative standard deviation, calculated from the peak heights obtained. The calibration graph preconcentration method for cobalt was linear with a correlation coefficient of 0.9994 from approximately 0.25 mg/L up to at least 100 mg/L. The method was successfully applied to the determination of cobalt in human urine samples.
Archivos asociados
Licencia
Identificadores
Colecciones
Articulos(INQUISAL)
Articulos de INST. DE QUIMICA DE SAN LUIS
Articulos de INST. DE QUIMICA DE SAN LUIS
Citación
Farias, Gustavo Marcelo; Wuilloud, Rodolfo German; Moyano, Susana; Gasquez, José Antonio; Olsina, Roberto Antonio; et al.; Determination of cobalt in urine by FI-ICP-AES with online preconcentration; Preston Publ Inc; Journal Of Analytical Toxicology; 26; 6; 1-9-2002; 360-364
Compartir
Altmétricas