Mostrar el registro sencillo del ítem

dc.contributor.author
Talio, María Carolina  
dc.contributor.author
Alesso, Magdalena  
dc.contributor.author
Acosta, Maria Gimena  
dc.contributor.author
Acosta, Mariano  
dc.contributor.author
Fernandez, Liliana Patricia  
dc.date.available
2016-05-20T20:38:06Z  
dc.date.issued
2014-09  
dc.identifier.citation
Talio, María Carolina; Alesso, Magdalena; Acosta, Maria Gimena; Acosta, Mariano; Fernandez, Liliana Patricia; Sequential determination of lead and cobalt in tap water and foods samples by fluorescence; Elsevier; Talanta; 127; 9-2014; 244-249  
dc.identifier.issn
0039-9140  
dc.identifier.uri
http://hdl.handle.net/11336/5763  
dc.description.abstract
In this work, a new procedure was developed for the separation and preconcentration of lead(II) and cobalt(II) in several water and foods samples. Complexes of metal ions with 8-hydroxyquinolein (8-HQ) were formed in aqueous solution. The proposed methodology is based on the preconcentration/separation of Pb(II) by solid-phase extraction using paper filter, followed by spectrofluorimetric determination of both metals, on the solid support and the filtered aqueous solution, respectively. The solid surfase fluorescence determination was carried out at λem = 455 nm (λex = 385 nm) for Pb(II)-8-HQ complex and the fluorescence of Co(II)-8-HQ was determined in aqueous solution using λem = 355 nm (λex = 225 nm). The calibration graphs are linears in the range 0.14 to 8.03 104 μg L−1 and 7.3× 10−2 to 4.12 × 103 μg L−1, for Pb(II) and Co(II), respectively, with a detection limit of 4.3 10−2 and 2.19 10−2 μg L−1 (S/N = 3). The developed methodology showed good sensitivity and adequate selectivity and it was successfully applied to the determination of trace amounts of lead and cobalt in tap waters belonging of different regions of Argentina and foods samples (milk powder, express coffee, cocoa powder) with satisfactory results. The new methodology was validated by electrothermal atomic absorption spectroscopy with adequate agreement. The proposed methodology represents a novel application of fluorescence to Pb(II) and Co(II) quantification with sensitivity and accuracy similar to atomic spectroscopies.  
dc.format
application/pdf  
dc.language.iso
eng  
dc.publisher
Elsevier  
dc.rights
info:eu-repo/semantics/openAccess  
dc.rights.uri
https://creativecommons.org/licenses/by-nc-nd/2.5/ar/  
dc.subject
Lead  
dc.subject
Cobalt  
dc.subject
8-Hydroxyquinoline  
dc.subject
Luminescence  
dc.subject
Tap Water  
dc.subject
Food  
dc.subject.classification
Química Analítica  
dc.subject.classification
Ciencias Químicas  
dc.subject.classification
CIENCIAS NATURALES Y EXACTAS  
dc.title
Sequential determination of lead and cobalt in tap water and foods samples by fluorescence  
dc.type
info:eu-repo/semantics/article  
dc.type
info:ar-repo/semantics/artículo  
dc.type
info:eu-repo/semantics/publishedVersion  
dc.date.updated
2016-05-16T20:16:35Z  
dc.journal.volume
127  
dc.journal.pagination
244-249  
dc.journal.pais
Países Bajos  
dc.journal.ciudad
Amsterdam  
dc.description.fil
Fil: Talio, María Carolina. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico San Luis. Instituto de Química de San Luis; Argentina  
dc.description.fil
Fil: Alesso, Magdalena. Universidad Nacional de San Luis. Facultad de Quimica, Bioquimica y Farmacia. Departamento de Quimica. Area de Quimica Analitica; Argentina  
dc.description.fil
Fil: Acosta, Maria Gimena. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico San Luis. Instituto de Química de San Luis; Argentina  
dc.description.fil
Fil: Acosta, Mariano. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico San Luis. Instituto de Química de San Luis; Argentina  
dc.description.fil
Fil: Fernandez, Liliana Patricia. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico San Luis. Instituto de Química de San Luis; Argentina. Universidad Nacional de San Luis. Facultad de Quimica, Bioquimica y Farmacia. Departamento de Quimica. Area de Quimica Analitica; Argentina  
dc.journal.title
Talanta  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/url/http://www.sciencedirect.com/science/article/pii/S0039914014002987  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/doi/10.1016/j.talanta.2014.04.020  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/doi/http://dx.doi.org/10.1016/j.talanta.2014.04.020