Mostrar el registro sencillo del ítem

dc.contributor.author
El Batti, Salma  
dc.contributor.author
Casciaro, Mariano Ezequiel  
dc.contributor.author
Alsac, Jean Marc  
dc.contributor.author
Latremouille, Christian  
dc.contributor.author
Julia, Pierre  
dc.contributor.author
Mousseaux, Elie  
dc.contributor.author
Craiem, Damian  
dc.date.available
2022-07-25T10:26:47Z  
dc.date.issued
2018-03  
dc.identifier.citation
El Batti, Salma; Casciaro, Mariano Ezequiel; Alsac, Jean Marc; Latremouille, Christian; Julia, Pierre; et al.; Abdominal Aortic Aneurysm Volumetric Evaluation During Mid-term Follow-Up After Endovascular Sealing Using the Nellix™ Device; Springer; Cardiovascular Engineering and Technology; 10; 1; 3-2018; 22-31  
dc.identifier.issn
1869-408X  
dc.identifier.uri
http://hdl.handle.net/11336/162964  
dc.description.abstract
Objectives: To analyze the volumetric evolution of abdominal aortic aneurysms after endovascular sealing (EVAS) with the Nellix™ device during follow-up. Methods: Patients who underwent elective EVAS in our institution in 2014 and 2015 were retrospectively reviewed. Preoperative, postoperative and 1-year scans were processed. A custom software was conceived to assess semi-automated measurements of the aneurysm sac and the endograft sizes including volume, maximum diameter, sectional area and perimeter. Thrombus volume, aneurysm length, mean distance between the stents inside the polymer-filled sacs and endograft migration were also estimated. Manual maximum diameters were measured for comparison. Inter and intra-observer variability of the proposed semi-automated method was evaluated. Results: Pre-EVAS, post-EVAS and last follow-up scans of 12 patients were finally analyzed during a mean follow-up of 17 ± 5 months. No endograft migration or endoleak were detected. During follow-up, aneurysm volume and perimeter slightly increased compared to post-EVAS scans (+ 1 and + 5%, respectively, p < 0.05). A systematic 6% enlargement of the endobag volume was also observed (range 1–15 mL, p < 0.001). Endobag maximum diameter, area and perimeter increased 4, 8, and 8%, respectively (all p < 0.01). Mean plane-by-plane distance between stents increased 4% (p < 0.05). Mean thrombus volume did not change during follow-up, although a high variability was observed. Aneurysm and thrombus volume changes were highly correlated (r = 0.93, p < 0.001). No associations were observed between aneurysm and endobag volume changes. Intra- and inter-observer variability was below 1.7 and 2.4% for diameter and volume measurements, respectively. The automated measurements of post-EVAS aneurysm diameter and volume were higher than preoperative (p < 0.05). Maximum diameters measured manually did not differ between scans. Conclusion: Small aneurysm volume enlargement detected during a mid-term follow-up was associated with thrombus size change, whereas systematic endograft expansion resulted independent from the aortic growth. Volumetric measurements using a semi-automated method could quantify small changes in aneurysm, endograft and thrombus sizes not detected by manually defined maximal diameters.  
dc.format
application/pdf  
dc.language.iso
eng  
dc.publisher
Springer  
dc.rights
info:eu-repo/semantics/openAccess  
dc.rights.uri
https://creativecommons.org/licenses/by-nc-sa/2.5/ar/  
dc.subject
ABDOMINAL AORTIC ANEURYSM  
dc.subject
ANEURYSM SIZING  
dc.subject
ANEURYSM VOLUME  
dc.subject
ENDOGRAFT FOLLOW-UP  
dc.subject
ENDOVASCULAR SEALING SYSTEM  
dc.subject.classification
Ingeniería Médica  
dc.subject.classification
Ingeniería Médica  
dc.subject.classification
INGENIERÍAS Y TECNOLOGÍAS  
dc.title
Abdominal Aortic Aneurysm Volumetric Evaluation During Mid-term Follow-Up After Endovascular Sealing Using the Nellix™ Device  
dc.type
info:eu-repo/semantics/article  
dc.type
info:ar-repo/semantics/artículo  
dc.type
info:eu-repo/semantics/publishedVersion  
dc.date.updated
2022-06-21T19:24:25Z  
dc.identifier.eissn
1869-4098  
dc.journal.volume
10  
dc.journal.number
1  
dc.journal.pagination
22-31  
dc.journal.pais
Estados Unidos  
dc.journal.ciudad
Nueva York  
dc.description.fil
Fil: El Batti, Salma. Hopital Europeen Georges Pompidou; Francia  
dc.description.fil
Fil: Casciaro, Mariano Ezequiel. Consejo Nacional de Investigaciones Científicas y Técnicas. Oficina de Coordinación Administrativa Houssay. Instituto de Medicina Traslacional, Trasplante y Bioingeniería. Fundación Favaloro. Instituto de Medicina Traslacional, Trasplante y Bioingeniería; Argentina  
dc.description.fil
Fil: Alsac, Jean Marc. Hopital Europeen Georges Pompidou; Francia  
dc.description.fil
Fil: Latremouille, Christian. Hopital Europeen Georges Pompidou; Francia  
dc.description.fil
Fil: Julia, Pierre. Hopital Europeen Georges Pompidou; Francia  
dc.description.fil
Fil: Mousseaux, Elie. Hopital Europeen Georges Pompidou; Francia  
dc.description.fil
Fil: Craiem, Damian. Consejo Nacional de Investigaciones Científicas y Técnicas. Oficina de Coordinación Administrativa Houssay. Instituto de Medicina Traslacional, Trasplante y Bioingeniería. Fundación Favaloro. Instituto de Medicina Traslacional, Trasplante y Bioingeniería; Argentina  
dc.journal.title
Cardiovascular Engineering and Technology  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/url/http://link.springer.com/10.1007/s13239-018-00380-4  
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/doi/http://dx.doi.org/10.1007/s13239-018-00380-4