Artículo
Characterisation and simulation of an active microvalve for glaucoma
Fecha de publicación:
08/2011
Editorial:
Taylor & Francis Ltd
Revista:
Computer Methods In Biomechanics And Biomedical Engineering
ISSN:
1025-5842
Idioma:
Inglés
Tipo de recurso:
Artículo publicado
Clasificación temática:
Resumen
Glaucoma drainage device (GDD) has the potential to eliminate hypotony
but still suffers from poor flow control and fibrosis. The ideal shunt
should change its hydraulic resistance to achieve the desired
intraocular pressure (IOP). In this study, the characterisation of a
preliminary design of a new GDD is presented. This is activated by means
of a diaphragm, which is actuated by conducting polymers. The valve can
be manufactured employing microelectromechanical system technology by
soft lithography. The characterisation process is performed by numerical
simulation using the finite element method, considering the coupling
between the fluid and the structure (diaphragm) obtaining the hydraulic
resistance for several positions of the diaphragm. To analyse the
hydraulic system of the microvalve implanted in a human eye, an
equivalent circuit model was used. The parameters of the equivalent
circuit model were obtained from numerical simulation. The hydraulic
resistance of the designed GDD varies in the range of
13.08–0.36 mmHg min/μl compared with 3.38–0.43 mmHg min/μl for the Ahmed
valve. The maximum displacement of the diaphragm in the vertical
direction is 18.9 μm, and the strain in the plane is 2%. The proposed
preliminary design allows to control the IOP by varying the hydraulic
resistance in a greater range than the existing passive valves, and the
numerical simulation facilitates the characterisation and the
improvement of the design before its construction, reducing time and
costs.
Palabras clave:
Mecánica de Fluidos Computacional
Archivos asociados
Licencia
Identificadores
Colecciones
Articulos(INTEC)
Articulos de INST.DE DES.TECNOL.PARA LA IND.QUIMICA (I)
Articulos de INST.DE DES.TECNOL.PARA LA IND.QUIMICA (I)
Citación
Sassetti, Fernando; Guarnieri, Fabio Ariel; Garelli, Luciano; Storti, Mario Alberto; Characterisation and simulation of an active microvalve for glaucoma
; Taylor & Francis Ltd; Computer Methods In Biomechanics And Biomedical Engineering; 8-2011
Compartir