dc.description.abstract
Background The novel signal processing techniques have allowed and improved the use of vectorcardiography (VCG) to diagnose and characterize myocardial ischemia. Herein, we studied vectorcardiographic dynamic changes of ventricular repolarization in 80 patients before (control) and during Percutaneous Transluminal Coronary Angioplasty (PTCA). Methods We propose four vectorcardiographic ST-T parameters, i.e., (a) ST Vector Magnitude Area (aSTVM); (b) T-wave Vector Magnitude Area (aTVM); (c) ST-T Vector Magnitude Difference (ST-TVD), and (d) T-wave Vector Magnitude Difference (TVD). For comparison, the conventional ST-Change Vector Magnitude (STCVM) and Spatial Ventricular Gradient (SVG) were also calculated. Results Our results indicate that several vectorcardiographic parameters show significant differences (p-value<0.05) before starting and during PTCA. Statistical minute-by-minute PTCA comparison against the control situation showed that ischemic monitoring reached a sensitivity=90.5% and a specificity=92.6% at the 5th minute of the PTCA, when aSTVM and ST-TVD were used as classifiers. Conclusions We conclude that the sensitivity and specificity for acute ischemia monitoring could be increased with the use of only two vectorcardiographic parameters. Hence, the proposed technique based on vectorcardiography could be used in addition to the conventional ST-T analysis for better monitoring of ischemic patients.
dc.format
application/pdf
dc.language.iso
eng
dc.publisher
Elsevier
dc.rights
info:eu-repo/semantics/openAccess
dc.rights.uri
https://creativecommons.org/licenses/by-nc-nd/2.5/ar/
dc.subject
St-T
dc.subject
Vcg
dc.subject
Ptca
dc.subject
Digital Signal Processing
dc.subject.classification
Otras Ingenierías y Tecnologías
dc.subject.classification
Otras Ingenierías y Tecnologías
dc.subject.classification
INGENIERÍAS Y TECNOLOGÍAS
dc.title
Novel technique for ST-T interval characterization in patients with acute myocardial ischemia
dc.type
info:eu-repo/semantics/article
dc.type
info:ar-repo/semantics/artículo
dc.type
info:eu-repo/semantics/publishedVersion
dc.date.updated
2016-03-21T18:28:54Z
dc.journal.volume
50
dc.journal.pagination
49-55
dc.journal.pais
Países Bajos
dc.journal.ciudad
Amsterdam
dc.description.fil
Fil: Correa, Raúl Julián. Universidad Nacional de San Juan. Facultad de Ingenieria. Departamento de Electronica y Automatica. Gabinete de Tecnologia Medica; Argentina
dc.description.fil
Fil: Arini, Pedro David. Consejo Nacional de Investigaciones Científicas y Técnicas. Oficina de Coordinación Administrativa Saavedra 15. Instituto Argentino de Matemática; Argentina. Universidad de Buenos Aires. Facultad de Ingeniería; Argentina
dc.description.fil
Fil: Correa, Lorena. Universidad Nacional de San Juan. Facultad de Ingenieria. Departamento de Electronica y Automatica. Gabinete de Tecnologia Medica; Argentina
dc.description.fil
Fil: Valentinuzzi, Max. Universidad de Buenos Aires. Facultad de Ingenieria. Instituto de Ingeniería Biomédica; Argentina
dc.description.fil
Fil: Laciar Leber, Eric. Universidad Nacional de San Juan. Facultad de Ingenieria. Departamento de Electronica y Automatica. Gabinete de Tecnologia Medica; Argentina. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina
dc.journal.title
Computers In Biology And Medicine
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/url/http://www.sciencedirect.com/science/article/pii/S001048251400095X
dc.relation.alternativeid
info:eu-repo/semantics/altIdentifier/doi/http://dx.doi.org/10.1016/j.compbiomed.2014.04.009
Archivos asociados